Image quality and dose in lumbar spine examinations: Results of a 5 year quality control programme following the European quality criteria trial

被引:19
作者
Vano, E
Oliete, S
Gonzalez, L
Guibelalde, E
Valasco, A
Fernandez, JM
机构
[1] UNIV MADRID,SAN CARLOS HOSP,DEPT MED PHYS,E-28040 MADRID,SPAIN
[2] INT RUBER HOSP,MADRID,SPAIN
关键词
D O I
10.1259/0007-1285-68-816-1332
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The large dose values found for lumbar spine examinations in a centre participating in the European quality criteria trial have been investigated within a 5 year quality control programme. Actions focused mainly on optimizing the focus-to-film distance, tube potential (kV), film optical density and X-ray beam filtration. These actions lead to overall dose reductions of up to 75% in lumbo-sacral joint examinations and prove the need for a regular survey of patient skin doses.
引用
收藏
页码:1332 / 1335
页数:4
相关论文
共 9 条
[1]  
MACCIA C, 1995, RADIAT PROT DOSIM, V57, P111
[2]  
MACCIA C, 1991, CECXII22193 DOC
[3]  
MCNEIL EA, 1995, RADIAT PROT DOSIM, V57, P423
[4]   CEC QUALITY CRITERIA FOR DIAGNOSTIC RADIOGRAPHIC IMAGES - BASIC CONCEPTS [J].
MOORES, BM .
RADIATION PROTECTION DOSIMETRY, 1995, 57 (1-4) :105-110
[5]  
MORAN B, 1995, RADIAT PROT DOSIM, V57, P423
[6]   ENTRANCE DOSES DURING LATERAL LUMBAR SPINE AND ANTERO-POSTERIOR ABDOMEN EXAMINATIONS - GENERATOR WAVE-FORM DEPENDENCE [J].
MORRISON, GD ;
UNDERWOOD, AC .
BRITISH JOURNAL OF RADIOLOGY, 1995, 68 (809) :491-494
[7]  
VANO E, 1992, RADIOLOGIA, V34, P27
[8]  
1991, XII17390 COMM EUR CO
[9]  
1982, METHODS SENSITOMETRY