MRA STUDIES OF ARTERIAL-STENOSIS - IMPROVEMENTS BY DIASTOLIC ACQUISITION

被引:28
作者
SALONER, D
SELBY, K
ANDERSON, CM
机构
[1] Department of Radiology, VA Medical Center, University of California San Francisco, San Francisco, California
关键词
MRA; ARTERIAL STENOSIS; DIASTOLIC ACQUISITION; FLOW PULSATILITY;
D O I
10.1002/mrm.1910310213
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Cardiac-phase-specific data acquisition is used to reduce signal loss in MR Angiography resulting from disturbed flow. RF pulses are delivered continuously throughout the cardiac cycle, but incrementation of phase-encoding gradients and data storage are enabled only during the chosen part of the cycle. Studies in a stenotic pulsatile flow phantom demonstrate that poststenotic signal loss is primarily determined by the mean flow velocity, and is not appreciably affected by acceleration or deceleration of the mean flow rate. The signal loss is least in diastole. In vivo studies in patients with carotid artery disease show that data acquisition in diastole reduces the apparent degree and extent of carotid bifurcation stenosis and provides a crisper definition of the vascular lumen. The additional time required for cardiac-phase-specific acquisition can be reduced by gating only the lower-order phase-encoding lines while retaining acceptable image quality.
引用
收藏
页码:196 / 203
页数:8
相关论文
共 20 条
[1]  
ANDERSON CM, 1992, AM J NEURORADIOL, V13, P989
[2]   CINEANGIOGRAPHY OF THE HEART IN A SINGLE BREATH HOLD WITH A SEGMENTED TURBOFLASH SEQUENCE [J].
ATKINSON, DJ ;
EDELMAN, RR .
RADIOLOGY, 1991, 178 (02) :357-360
[3]   IMAGING OF THE RENAL-ARTERIES - VALUE OF MR ANGIOGRAPHY [J].
DEBATIN, JF ;
SPRITZER, CE ;
GRIST, TM ;
BEAM, C ;
SVETKEY, LP ;
NEWMAN, GE ;
SOSTMAN, HD .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1991, 157 (05) :981-990
[4]   MR ANGIOGRAPHY WITH PULSATILE FLOW [J].
DEGRAAF, RG ;
GROEN, JP .
MAGNETIC RESONANCE IMAGING, 1992, 10 (01) :25-34
[5]  
FRANCK A, 1993, J MAGN RESON IMAGI P, V3, pS31
[6]   ELIMINATION OF OBLIQUE FLOW ARTIFACTS IN MAGNETIC-RESONANCE-IMAGING [J].
FRANK, LR ;
CRAWLEY, AP ;
BUXTON, RB .
MAGNETIC RESONANCE IN MEDICINE, 1992, 25 (02) :299-307
[7]   THE SALIENT FEATURES OF MR ANGIOGRAPHY [J].
HAACKE, EM ;
MASARYK, TJ .
RADIOLOGY, 1989, 173 (03) :611-612
[8]   ABDOMINAL-AORTA AND RENAL-ARTERY STENOSIS - EVALUATION WITH MR ANGIOGRAPHY [J].
KIM, D ;
EDELMAN, RR ;
KENT, KC ;
PORTER, DH ;
SKILLMAN, JJ .
RADIOLOGY, 1990, 174 (03) :727-731
[9]   MR IMAGING OF POSTSTENOTIC FLOW PHENOMENA - EXPERIMENTAL STUDIES [J].
KRUG, B ;
KUGEL, H ;
FRIEDMANN, G ;
BUNKE, J ;
VANDIJK, P ;
SCHMIDT, R ;
HIRCHE, HJ .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1991, 1 (05) :585-591
[10]  
MCDONALD DA, 1974, BLOOD FLOW ARTERIES