Clinical implementation of dynamic intensity-modulated radiotherapy: Dosimetric aspects and initial experience

被引:4
作者
Sivakumar, S. S. [1 ]
Krishnamurthy, K. [1 ]
Davis, C. A. [1 ]
Ravichandran, R. [1 ]
Kannadhasan, S. [1 ]
Biunkumar, J. P. [1 ]
El Ghamrawy, Kamal [1 ]
机构
[1] Royal Hosp, Dept Radiat Oncol, Natl Oncol Ctr, Post Box 1331,PC 111, Muscat, Oman
关键词
Dose verification; intensity-modulated radiotherapy; millennium multi-leaf collimator; quality assurance test; sliding window;
D O I
10.4103/0971-6203.41195
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This paper describes the initial experience of quality assurance (QA) tests performed on the millennium multi-leaf collimator (mMLC) for clinical implementation of intensity-modulated radiotherapy (IMRT) using sliding window technique. The various QA tests verified the mechanical and dosimetric stability of the mMLC of linear accelerator when operated in dynamic mode (dMLC). The mechanical QA tests also verified the positional accuracy and kinetic properties of the dMLC. The stability of dMLC was analyzed qualitatively and quantitatively using radiographic film and Omnipro IMRT software. The output stability, variation in output for different sweeping gap widths, and dosimetric leaf separation were measured. Dose delivery with IMRT was verified against the dose computed by the treatment planning system (TPS). Monitor units (MUs) calculated by the planning system for the IMRT were cross-checked with independent commercial dose management software. Visual inspection and qualitative analysis showed that the leaf positioning accuracy was well within the acceptable limits. Dosimetric QA tests confirmed the dosimetric stability of the mMLC in dynamic mode. The verification of MUs using commercial software confirmed the reliability of the IMRT planning system for dose computation. The dosimetric measurements validated the fractional dose delivery.
引用
收藏
页码:64 / 71
页数:8
相关论文
共 12 条
  • [1] [Anonymous], QA TEST PATT PROC P
  • [2] [Anonymous], DYN FIL VIEW REF GUI
  • [3] Ayyangar KM, 2005, INT C MED PHYS RAD S, V30, P81
  • [4] BOHSUNG J, 2001, P VAR US M SARD VAR
  • [5] Testing of dynamic multileaf collimation
    Chui, CS
    Spirou, S
    LoSasso, T
    [J]. MEDICAL PHYSICS, 1996, 23 (05) : 635 - 641
  • [6] Haslam J J, 2003, J Appl Clin Med Phys, V4, P224, DOI 10.1120/1.1590611
  • [7] A monitor unit verification calculation in intensity modulated radiotherapy as a dosimetry quality assurance
    Kung, JH
    Chen, GTY
    Kuchnir, FK
    [J]. MEDICAL PHYSICS, 2000, 27 (10) : 2226 - 2230
  • [8] Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy
    LoSasso, T
    Chui, CS
    Ling, CC
    [J]. MEDICAL PHYSICS, 1998, 25 (10) : 1919 - 1927
  • [9] Comprehensive quality assurance for the delivery of intensity modulated radiotherapy with a multileaf collimator used in the dynamic mode
    LoSasso, T
    Chui, CS
    Ling, CC
    [J]. MEDICAL PHYSICS, 2001, 28 (11) : 2209 - 2219
  • [10] Implementation of IMRT in the radiotherapy department 10.1259/bjr/19737738
    McNair, HA
    Adams, EJ
    Clark, CH
    Miles, EA
    Nutting, CM
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2003, 76 (912) : 850 - 856