A DIFFERENTIAL PARAMAGNETIC SENSOR FOR BREATH-BY-BREATH OXIMETRY

被引:27
|
作者
MERILAINEN, PT
机构
[1] Datex Instrumentarium Corporation, Helsinki, SF-00101
来源
JOURNAL OF CLINICAL MONITORING | 1990年 / 6卷 / 01期
关键词
differential oxygen measurement; Measurement techniques: oximetry; breath-by-breath; Monitoring: oxygen; Oxygen:; measurement; sensors;
D O I
10.1007/BF02832185
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The operating principles of a fast paramagnetic oxygen sensor are described in detail. Some historical background of the paramagnetic effect in oxygen measurement is given as well. This sensor measures the difference between oxygen partial pressures in two gases as an alternating pressure signal being generated by the interaction of oxygen molecules with a switched magnetic field. Factors affecting the ideal performance of the sensor are analyzed. The main sources of interference are external pressure and vibration, and asymmetry in the transfer of the pressure signal to the microphone. Finally, the advantages of clinical breath-by-breath oximetry are briefly discussed. © 1990 Little, Brown and Company.
引用
收藏
页码:65 / 73
页数:9
相关论文
共 50 条
  • [1] Breath-by-breath measurement of oxygen using a compact optical sensor
    Burke, Conor S.
    Moore, John P.
    Wencel, Dorota
    McEvoy, Aisling K.
    MacCraith, Brian D.
    JOURNAL OF BIOMEDICAL OPTICS, 2008, 13 (01)
  • [2] Exercise gas exchange, breath-by-breath
    Wasserman, K
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2002, 165 (03) : 325 - 326
  • [3] BREATH-BY-BREATH ASSESSMENT OF RESPONSE TO BRONCHODILATOR
    SOBOL, BJ
    REED, A
    WALDIE, J
    JOURNAL OF ALLERGY AND CLINICAL IMMUNOLOGY, 1973, 52 (06) : 358 - 362
  • [4] Breath-by-breath measurement of carbon dioxide using a plastic film optical sensor
    Mills, A
    Lepre, A
    Wild, L
    SENSORS AND ACTUATORS B-CHEMICAL, 1997, 39 (1-3) : 419 - 425
  • [5] BREATH-BY-BREATH ALVEOLAR GAS-EXCHANGE
    GIEZENDANNER, D
    CERRETELLI, P
    DIPRAMPERO, PE
    JOURNAL OF APPLIED PHYSIOLOGY, 1983, 55 (02) : 583 - 590
  • [6] BREATH-BY-BREATH DETERMINATION OF INSPIRATORY OCCLUSION PRESSURE
    LARSSON, H
    HELLSTROM, LG
    LINNARSSON, D
    CLINICAL PHYSIOLOGY, 1993, 13 (02): : 133 - 142
  • [7] Development of a compact optical sensor for real-time, breath-by-breath detection of oxygen
    Burke, C. S.
    Moore, J. P.
    Wencel, D.
    MacCraith, B. D.
    JOURNAL OF BREATH RESEARCH, 2008, 2 (03)
  • [8] PULSED TITRATION SENSORS .1. A BREATH-BY-BREATH CO2 SENSOR
    ALBERY, WJ
    CLARK, D
    DRUMMOND, HJJ
    COOMBS, AJM
    YOUNG, WK
    HAHN, CEW
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1992, 340 (1-2): : 99 - 110
  • [9] BREATH-BY-BREATH MEASUREMENT OF ALVEOLAR GAS-EXCHANGE
    LAFORTUNA, CL
    DIPRAMPERO, PE
    JOURNAL OF PHYSIOLOGY-LONDON, 1986, 378 : P67 - P67
  • [10] BREATH-BY-BREATH SAMPLING OF END-EXPIRATORY GAS
    LAMBERTSEN, CJ
    BENJAMIN, JM
    JOURNAL OF APPLIED PHYSIOLOGY, 1959, 14 (05) : 711 - 716