CRYOGENIC COOLING OF MONOCHROMATORS

被引:59
作者
MAROT, G
ROSSAT, M
FREUND, A
JOKSCH, S
KAWATA, H
ZHANG, L
ZIEGLER, E
BERMAN, L
CHAPMAN, D
HASTINGS, JB
IAROCCI, M
机构
[1] DESY,HASYLAB,W-2000 HAMBURG 52,GERMANY
[2] BROOKHAVEN NATL LAB,NATL SYNCHROTRON LIGHT SOURCE,UPTON,NY 11973
关键词
D O I
10.1063/1.1142736
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In this paper, we report on the design and x-ray diffraction properties of cryocooled silicon single crystals exposed to the following wiggler beams: power density: 150 W/mm2, total power: 75 W and power density: 0.5 W/mm2, total power: 100 W. First, thermomechanical and engineering aspects of low-temperature crystal cooling are discussed, leading to two basic cooling geometries: internal cooling and side cooling. Experimental tests of both these cooling schemes at NSLS on beam lines X25 and X17 are then described and discussed. Finally, engineering problems related to the integration of cryogenic cooling on ESRF beam lines are presented.
引用
收藏
页码:477 / 480
页数:4
相关论文
共 12 条
[1]  
BERMAN LE, 1992, REV SCI INSTRUM, V63
[2]   THE POTENTIAL OF CRYOGENIC SILICON AND GERMANIUM X-RAY MONOCHROMATORS FOR USE WITH LARGE SYNCHROTRON HEAT LOADS [J].
BILDERBACK, DH .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1986, 246 (1-3) :434-436
[3]  
FREUND AK, 1990, OPT ENG, V29, P928, DOI 10.1117/12.55678
[4]   SURFACE-TEMPERATURE AND SLOPE ERRORS ON PLANE MIRRORS EXPOSED TO HIGH-POWER SYNCHROTRON RADIATION [J].
JARK, W ;
MOURIKIS, S ;
JOKSCH, S ;
SAILE, V .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 291 (1-2) :319-324
[5]  
JOKSCH S, 1990, UNPUB NUCL INSTRUM M, P583
[6]  
SCHILDKAMP W, 1990, 1989 WORKSH HIGH HEA
[7]  
Touloukian Y.S., 1970, THERMOPHYSICAL PROPE, V1st
[8]  
ZHANG L, 1990, 1989 WORKSH HIGH HEA
[9]  
1989, 1988 WORKSH COOL XRA
[10]  
1976, GAS ENCY AIR LIQUIDE