COMPLETE MASS-SPECTROMETRY MASS-SPECTROMETRY DATA FIELD ACQUISITION ON THE CHROMATOGRAPHIC TIME SCALE BY TIME-RESOLVED ION MOMENTUM SPECTROMETRY WITH TIME-ARRAY DETECTION

被引:3
作者
ECKENRODE, BA
WATSON, JT
ENKE, CG
HOLLAND, JF
机构
[1] MICHIGAN STATE UNIV, DEPT CHEM, E LANSING, MI 48824 USA
[2] MICHIGAN STATE UNIV, DEPT BIOCHEM, E LANSING, MI 48824 USA
关键词
D O I
10.1021/ac00213a003
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Time-resolved ion momentum spectrometry (TRIMS) is combined with time-array detection (TAD) to provide the capability for rapid acquisition of complete mass spectrometry/ mass spectrometry (MS/MS) data fields. Observation of all arrival times for all values of the magnetic field strength produces a two-dimensional field from which all MS/MS scans can be obtained. The complete fragmentation map can now be obtained for each component in a mixture, separated by chromatography. Evaluation of the TRIMS-TAD Instrument with regard to data quality, integrity, and speed is performed under raw data acquisition and real-time data reduction conditions. A 5-s acquisition of the MS/MS data field for caffeine present in an aspirin tablet demonstrates the gas chromatography (GC)-MS/MS capability of TRIMS-TAD. © 1990, American Chemical Society. All rights reserved.
引用
收藏
页码:1362 / 1367
页数:6
相关论文
共 33 条
[1]   DESIGN AND PERFORMANCE OF A MASS ANALYZED ION KINETIC-ENERGY (MIKE) SPECTROMETER [J].
BEYNON, JH ;
COOKS, RG ;
AMY, JW ;
BAITINGE.WE ;
RIDLEY, TY .
ANALYTICAL CHEMISTRY, 1973, 45 (12) :1023-+
[2]   COLLISION-INDUCED DISSOCIATION WITH FOURIER-TRANSFORM MASS-SPECTROMETRY [J].
CODY, RB ;
BURNIER, RC ;
FREISER, BS .
ANALYTICAL CHEMISTRY, 1982, 54 (01) :96-101
[3]   FOURIER-TRANSFORM ION-CYCLOTRON RESONANCE SPECTROSCOPY [J].
COMISAROW, MB ;
MARSHALL, AG .
CHEMICAL PHYSICS LETTERS, 1974, 25 (02) :282-283
[4]  
Cooks R. G., 1973, METASTABLE IONS
[5]  
COOKS RG, 1981, CHEM ENG NEWS 1130, P40
[6]  
Coutant J. E., 1972, International Journal of Mass Spectrometry and Ion Physics, Vvol8, P323, DOI 10.1016/0020-7381(72)83021-3
[7]   POST-SECTOR BEAM DEFLECTION IN TIME-RESOLVED ION MOMENTUM SPECTROMETRY (TRIMS) [J].
ECKENRODE, BA ;
WATSON, JT ;
ENKE, CG ;
HOLLAND, JF .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1988, 83 (1-2) :177-187
[8]   FLASH ADC TIME-OF-FLIGHT SPECTROMETER [J].
EGAN, PO ;
WOODLE, K .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1982, 53 (08) :1267-1269
[9]  
ENKE CG, 1987, ANAL CHEM, V59, pA1363
[10]  
ENKE CG, Patent No. 4490806