A novel method for salvage of malfunctioning peritoneal dialysis catheter

被引:2
作者
Beigi, Ali Akbar [1 ]
Marashi, Sayed Mahdi [2 ]
Asadabadi, Hojatollah Raji [1 ]
Sharifi, Ali [1 ]
Zarch, Zohre Nasiri [3 ]
机构
[1] Univ Tehran Med Sci, Bahrami Children Hosp, Dept Pediat Surg, Tehran, Iran
[2] Legal Med Org, Legal Med Res Ctr, Tehran, Iran
[3] Shahid Beheshti Univ, Fac Educ Sci & Psychol, Tehran, Iran
关键词
Catheter malfunction; catheter rescue; continuous ambulatory peritoneal dialysis; minimally invasive surgery; peritoneal dialysis;
D O I
10.4103/0974-7796.130646
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Context: Continuous ambulatory peritoneal dialysis (CAPD) has been widely used as an effective therapy in the management of patients with end-stage renal disease. Long-term use of CAPD needs methods with low incidence of catheter-related complications. Moreover, some complications may cause failure of fluid drainage and treatment interruption. Aims: We have innovated and studied a new minimal-invasive method of malfunctioning peritoneal catheter repair. Materials and Methods: Thirty-five patients agreed to undergo catheter rescue operation by this new method during 2004 and 2012. Under local anesthesia and light sedation, access to the abdominal cavity was made, the catheter and wrapped omentum grasped and the tip of catheter was released, debris were removed and the catheter was directed toward the pelvic floor with a finger guide. The patients were followed after catheter salvage up to the end of study (April 2012). PD catheter function restored to the normal level in 28 (80%) of patients, and PD was started 1-2 days after the procedure. Results: All patients had an uneventful recovery. PD catheter function was restored to the normal level in 28 (80%) patients, and PD was started 1-2 days after the procedure. Of these patients, 10 (35%) died of reasons unrelated to catheter or catheter complications; 7 (25%) were ultimately referred for kidney transplant; 8 (29%) continued PD up to the end of this study with no problem, and only 3 (11%) due to catheter complications. Catheter function did not restore to the normal level in seven patients (20%); however, six patients continued PD for 1-18 months with the catheter. Conclusions: Comparing the advantages and disadvantages of this method to the previous laparoscopically repaired catheter, we concluded that this new method is efficient, and is a suitable way for malfunctioning PD catheter salvage.
引用
收藏
页码:147 / 151
页数:5
相关论文
共 34 条
[21]   Laparoscopic tenckhoff catheter insertion: A prospective study of a new technique [J].
Poole, GH ;
Tervit, P .
AUSTRALIAN AND NEW ZEALAND JOURNAL OF SURGERY, 2000, 70 (05) :371-373
[22]  
Riella MC, 2005, CONTRIB NEPHROL, V149, P131, DOI 10.1159/000085476
[23]  
SCHAUER PR, 2000, MASTERY ENDOSCOPIC S, P22
[24]   Laparoscopic-assisted placement of peritoneal dialysis catheters: Implantation technique and results [J].
Schmidt, Sven C. ;
Pohle, Cosima ;
Langrehr, Jan M. ;
Schumacher, Guido ;
Jacob, Dietmar ;
Neuhaus, Peter .
JOURNAL OF LAPAROENDOSCOPIC & ADVANCED SURGICAL TECHNIQUES, 2007, 17 (05) :596-599
[25]   VALUE OF PERITONEOSCOPY FOR NONFUNCTIONING CONTINUOUS AMBULATORY PERITONEAL-DIALYSIS CATHETERS [J].
SMITH, DW ;
RANKIN, RA .
GASTROINTESTINAL ENDOSCOPY, 1989, 35 (02) :90-92
[26]  
Tiong H. Y., 2006, SMJ Singapore Medical Journal, V47, P707
[27]  
TISHER CC, 1994, ANN INTERN MED, V121, P62
[28]  
Uchiyama K, 2001, Hinyokika Kiyo, V47, P669
[29]  
*US REN DAT SYST, 2000, AM J KIDNEY DIS S, V36, pS127
[30]  
Wang JY, 1999, AM SURGEON, V65, P247