Accuracy of Intraoperative Computed Tomography-Based Navigation for Placement of Percutaneous Pedicle Screws

被引:15
作者
Eck, Jason C. [1 ]
Lange, Jeffrey [1 ]
Street, John [2 ]
Lapinsky, Anthony [1 ]
DiPaola, Christian P. [1 ]
机构
[1] Univ Massachusetts, Dept Orthoped Surg, 119 Belmont St, Worcester, MA 01564 USA
[2] Univ British Columbia, Dept Orthoped Surg, Vancouver, BC, Canada
关键词
minimally invasive spine surgery; percutaneous pedicle screw; navigation; accuracy; instrumentation;
D O I
10.1055/s-0033-1345037
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
MISS techniques have gained recent popularity. The proposed benefits of these techniques include reduced tissue trauma, reduced blood loss, less perioperative pain, and a quicker recovery and return to normal activities. The purpose of this study was to evaluate the accuracy of intraoperative computed tomography (CT)-based navigation for placement of percutaneous pedicle screws in a cadaveric model. Outcome measures included accuracy of screw placement. Two cadaveric specimens were utilized. CT images were obtained using an O-Arm (Medtronic, Memphis, Tennessee, United States) and were coupled to the Stealth navigation system (Medtronic). Computer navigation was used for placement of percutaneous pedicle screws. Screws were placed bilaterally from T5 to S1. Postinsertion CT scans were obtained. Pedicle breach was assessed and classified (I: none, II: < 2 mm, III: 2 to 4 mm, or IV: > 4 mm) with direction of breach. Thirty thoracic screws were placed with 3 (10%) medial breaches and 17 (56.7%) lateral breaches (grade III). Of 20 lumbar screws there were 0 medial breaches and 2 (10%) lateral breaches (1 grade III, 1 grade IV). Four sacral screws were placed without breaches. The real-time computer-aided navigation tool ("simulated screw") was limited in identifying a breach. Manipulation of the surgeon's hand or driver could change the orientation of the navigation tool without changing the screw trajectory. CT-based navigation for percutaneous pedicle screw placement appears safe for the lumbar spine. Lateral thoracic breaches appeared commonly but were not felt to be clinically significant. The 10% rate of medial thoracic breach was concerning, but definitive conclusions could not be made due to the small sample size.
引用
收藏
页码:103 / 107
页数:5
相关论文
共 11 条
[1]   Minimally Invasive Multilevel Percutaneous Correction and Fusion for Adult Lumbar Degenerative Scoliosis A Technique and Feasibility Study [J].
Anand, Neel ;
Baron, Eli M. ;
Thaiyananthan, Gowriharan ;
Khalsa, Kunwar ;
Goldstein, Theodore B. .
JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2008, 21 (07) :459-467
[2]  
Best Natalie M, 2009, Am J Orthop (Belle Mead NJ), V38, P387
[3]   Minimally invasive lumbar spinal fusion [J].
Eck, Jason C. ;
Hodges, Scott ;
Humphreys, S. Craig .
JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2007, 15 (06) :321-329
[4]   Accuracy of pedicle screw placement: a systematic review of prospective in vivo studies comparing free hand, fluoroscopy guidance and navigation techniques [J].
Gelalis, Ioannis D. ;
Paschos, Nikolaos K. ;
Pakos, Emilios E. ;
Politis, Angelos N. ;
Arnaoutoglou, Christina M. ;
Karageorgos, Athanasios C. ;
Ploumis, Avraam ;
Xenakis, Theodoros A. .
EUROPEAN SPINE JOURNAL, 2012, 21 (02) :247-255
[5]   Factors affecting the accurate placement of percutaneous pedicle screws during minimally invasive transforaminal lumbar interbody fusion [J].
Kim, Moon-Chan ;
Chung, Hung-Tae ;
Cho, Jae-Lim ;
Kim, Dong-Jun ;
Chung, Nam-Su .
EUROPEAN SPINE JOURNAL, 2011, 20 (10) :1635-1643
[6]   Radiation Exposure to the Surgeon During Percutaneous Pedicle Screw Placement [J].
Mroz, Thomas E. ;
Abdullah, Kalil G. ;
Steinmetz, Michael P. ;
Klineberg, Eric O. ;
Lieberman, Isador H. .
JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2011, 24 (04) :264-267
[7]   Comparison of the Percutaneous Screw Placement Precision of Isocentric C-arm 3-dimensional Fluoroscopy-navigated Pedicle Screw Implantation and Conventional Fluoroscopy Method With Minimally Invasive Surgery [J].
Nakashima, Hiroaki ;
Sato, Kouji ;
Ando, Toshihiro ;
Inoh, Hidefumi ;
Nakamura, Hiroshi .
JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2009, 22 (07) :468-472
[8]   Comparison of one-level posterior lumbar interbody fusion performed with a minimally invasive approach or a traditional open approach [J].
Park, Yung ;
Ha, Joong Won .
SPINE, 2007, 32 (05) :537-543
[9]   Clinical Accuracy of Computer-Assisted Two-Dimensional Fluoroscopy for the Percutaneous Placement of Lumbosacral Pedicle Screws [J].
Ravi, Bheeshma ;
Zahrai, Ali ;
Rampersaud, Raja .
SPINE, 2011, 36 (01) :84-91
[10]   The Accuracy of Intraoperative O-arm Images for the Assessment of Pedicle Screw Postion [J].
Santos, Edward R. G. ;
Ledonio, Charles G. ;
Castro, Carlos A. ;
Truong, Walter H. ;
Sembrano, Jonathan N. .
SPINE, 2012, 37 (02) :E119-E125