DETERMINATION OF CADMIUM IN WHOLE-BLOOD AND URINE BY ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY USING PALLADIUM-BASED MODIFIERS AND IN-SITU DECONTAMINATION

被引:0
作者
MOREIRA, FDR
CURTIUS, AJ
DECAMPOS, RC
机构
[1] PONTIFICIA UNIV CATOLICA RIO DE JANEIRO,DEPT QUIM,RIO JANEIRO,BRAZIL
[2] FDN OSVALDO CRUZ,RIO JANEIRO,BRAZIL
[3] UNIV FED SANTA CATARINA,DEPT QUIM,FLORIANOPOLIS,SC,BRAZIL
[4] UNIV FED RURAL RIO DE JANEIRO,DEPT QUIM,RIO JANEIRO,BRAZIL
关键词
CADMIUM; WHOLE BLOOD; URINE; ELECTROTHERMAL ATOMIC ABSORPTION SPECTROMETRY; IN SITU DECONTAMINATION; MODIFIER;
D O I
暂无
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The determination of Cd in whole blood and urine by electrothermal atomic absorption spectrometry using Pd-based modifiers and in situ decontamination was studied, The performance of Pd, Pd + NH4NO3, and Pd + Mg(NO3)(2) as modifiers were compared with respect to maximum pyrolysis and atomization temperatures, background attenuation, linear calibration range and matrix interference, Samples were diluted (1 + 4) with 0.1% Triton X-100 (blood) or 0.2% HNO3 (urine), and introduced onto the graphite tube platform already containing the in situ dried and decontaminated modifier, The influence of modifier volume and concentration, as well as that of the dilution ratio was investigated, The proposed procedure involves the use of Pd(NO3)(2) + Mg(NO3)(2) as modifier, a pyrolysis temperature of 700 degrees C and an in situ decontamination step of 1200 degrees C, The atomization temperature was 1700 degrees C. The detection limit (n = 10, k = 3) was 0.22 mu g l(-1) for both samples, Typical relative standard deviation values (n = 3) were below 3%. Using matrix-matched standards, good agreement was observed between experimental and reference values in the analysis of certified reference materials, Interlaboratory comparison data have confirmed the validity of the proposed analytical procedure.
引用
收藏
页码:947 / 950
页数:4
相关论文
共 10 条
[1]  
[Anonymous], 1988, INTRO MICROWAVE SAMP
[2]  
BLACK MM, 1986, J ANAL ATOM SPECTROM, V5, P393
[3]   INSITU REMOVAL OF CONTAMINATION FROM A PALLADIUM-MAGNESIUM CHEMICAL MODIFIER WITHIN AN ELECTROTHERMAL ATOMIZER FOR THE DETERMINATION OF CADMIUM IN SERUM [J].
BULSKA, E ;
GROBENSKI, Z ;
SCHLEMMER, G .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1990, 5 (03) :203-204
[4]  
CASTILLO JR, 1987, ATOM SPECTROSC, V8, P172
[5]   DIRECT DETERMINATION OF CADMIUM IN URINE BY ELECTROTHERMAL ATOMIZATION ATOMIC-ABSORPTION SPECTROMETRY [J].
HALLS, DJ ;
BLACK, MM ;
FELL, GS ;
OTTAWAY, JM .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1987, 2 (03) :305-309
[6]  
KNAPP G, 1992, DETERMINATION TRACE
[7]  
SUBRAMANIAN KS, 1986, PROG ANAL SPECTROSC, V9, P237
[8]  
SUBRAMANIAN KS, 1988, PROG ANAL SPECTROSC, V11, P511
[9]  
VANLOON JC, 1991, SELECTED METHODS TRA
[10]  
WATANABE T, 1982, TOXICOL LETT, V13, P231