AUTOMATED REAL-TIME DATA ACQUISITION AND ANALYSIS OF CARDIORESPIRATORY FUNCTION

被引:3
作者
MOORMAN, RC
MACKENZIE, CF
HO, GH
BARNAS, GM
WILSON, PD
MATJASKO, MJ
机构
[1] The Department of Anesthesiology, University of Maryland, Baltimore, 21201, Maryland
[2] The Department of Epidemiology and Preventive Medicine, University of Maryland, Baltimore, 21201, Maryland
来源
INTERNATIONAL JOURNAL OF CLINICAL MONITORING AND COMPUTING | 1991年 / 8卷 / 01期
关键词
MICROCOMPUTER; DATA COLLECTION; CARDIAC RESPIRATORY FUNCTION;
D O I
10.1007/BF02916093
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Microcomputer generation of an automated record without complexity or operator intervention is desirable in many circumstances. We developed a microcomputer system specifically designed for simplified automated collection of cardiorespiratory data in research and clinical environments. We tested the system during possible extreme clinical conditions by comparison with a patient simulator. Ranges used were heart rate of 35-182 beats per minute, systemic blood pressures of 65-147 mmHg and venous blood pressures of 14-37 mmHg, all with superimposed respiratory variation of 0-24 mmHg. We also tested multiple electrocardiographic dysrhythmias. The results showed that there were no clinically relevant differences in vascular pressures, heart rate, and other variables between computer processed and simulator values. Manually and computer recorded physiological variables were compared to simulator values and the results show that computer values were more accurate. The system was used routinely in 21 animal research experiments over a 4 month period employing a total of 270 collection periods. The file system integrity was tested and found to be satisfactory, even during power failures. Unlike other data collection systems this one (1) requires little or no operator intervention and training, (2) has been rigorously tested for accuracy using a wide variety of extreme patient conditions, (3) has had computer derived values measured against a standardized reference, (4) is reliable against external sources of computer failure, and (5) has screen and printout presentations with quick and easily understandable formats.
引用
收藏
页码:59 / 69
页数:11
相关论文
共 12 条
  • [1] Block F.E., Burton L.B., Rafal M.D., Et al., Two computer-based anesthetic monitors: the Duke automatic monitoring equipment (DAME) system and the microDAME, Journal of Clinical Monitoring, 1, pp. 30-51, (1985)
  • [2] Gravenstein J.S., Newbower R., Ream A.K., Smith N.T., The automated anesthesia record and alarm systems, pp. 3-8, (1987)
  • [3] Prakash O., van der Borden S.G., Meij S.H., Et al., A microcomputer based charting system for documentation of circulatory, respiratory and pharmacologic data during anesthesia, International Journal of Clinical Monitoring and Computing, 1, pp. 155-160, (1984)
  • [4] Ream A.K., Automating the recording and improving the presentation of the anesthesia, record, Journal of Clinical Monitoring, 5, pp. 270-283, (1989)
  • [5] Tsang P.S., Johnson W.W., Cognitive demands in automation, Aviation Space and Environmental Medicine, 60, pp. 130-135, (1989)
  • [6] Norman D.A., The problem's with automation: inappropriate feedback and interaction, not ‘over-automation’, Philosophical transactions of the Royal Society of London. Series B: Biological Sciences, 12, pp. 585-593, (1990)
  • [7] King P.H., Blanks S., Smith B.E., Operating room monitoring and automated record-keeping systems, Hospimedica, 49, pp. 39-47, (1988)
  • [8] Schwid H.A., Olson C., Wright P., Freund P.R., Microcomputer-based data acquisition system for clinical research, Journal of Clinical Monitoring, 6, pp. 141-146, (1990)
  • [9] Wong D.H., Tremper K.K., Stemmer E.A., Et al., Non-invasive cardiac output: Simultaneous comparison of two different methods with thermodilution, Anesthesiology, 72, pp. 784-792, (1990)
  • [10] Shabot M.M., LoBue M., Leyerle B.J., Decision support alerts for clinical laboratory and blood gas data, International Journal of clinical Monitoring and Computing, 7, pp. 27-31, (1990)