MICRAL-TEST (R) - A QUALITATIVE DIPSTICK TEST FOR MICROALBUMINURIA

被引:21
作者
SPOOREN, PFMJ
LEKKERKERKER, JFF
VERMES, I
机构
[1] Department of Internal Medicine and Clinical Chemistry, Medisch Spectrum Twente, Enschede
关键词
DIABETIC NEPHROPATHY; DIPSTICK TEST; MICROALBUMINURIA;
D O I
10.1016/0168-8227(92)90003-A
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Micro-albuminuria is a very sensitive predictor of the development of renal disease in insulin- and non-insulin-dependent diabetes mellitus. A reliable dipstick test for routine screening for micro-albuminuria is, therefore, desirable. Such a test has been developed by Boehringer Mannheim, Germany, and marketed as Micral-Test(R). It is an immunological slide-test with semi-quantitative properties. To evaluate its performance as a screening test we compared it with a turbidimetric immuno-assay. In 396 urine specimens from 132 patients, sensitivity was 91 % and specificity 96 % for a discriminating albumin level of 20 mg/l. Correlation with quantitative values was reasonable (r = 0.73). We also tested for micro-albuminuria, defined as mean albumin excretion rate of greater-than-or-equal-to 20 mug/min, determined with the turbidimetric immuno-assay in timed ovenight urines on three consecutive days, whereas Micral-Test was considered to be positive for micro-albuminuria if the albumin concentration in one of the three urine samples was greater-than-or-equal-to 20 mg/ml. In 132 patients, the sensitivity of Micral-Test was 82 % and the specificity 86 %. Albumin excretion rate in all false-negative results was < 50 mug/min. We therefore concluded that Micral-Test is a useful qualitative screening test for micro-albuminuria in diabetic patients.
引用
收藏
页码:83 / 87
页数:5
相关论文
共 12 条
  • [1] NEW SEMIQUANTITATIVE DIPSTICK TEST FOR MICROALBUMINURIA
    BANGSTAD, HJ
    TRY, K
    DAHLJORGENSEN, K
    HANSSEN, KF
    [J]. DIABETES CARE, 1991, 14 (11) : 1094 - 1097
  • [2] CASSUTO D, 1991, INT J OBESITY S1, V15, P37
  • [3] FELDTRASMUSSEN B, 1986, LANCET, V2, P1300
  • [4] TURBIDIMETRIC MEASUREMENT OF MICROALBUMINURIA
    HARMOINEN, A
    VUORINEN, P
    JOKELA, H
    [J]. CLINICA CHIMICA ACTA, 1987, 166 (01) : 85 - 89
  • [5] Jarrett R J, 1984, Diabet Med, V1, P17
  • [6] MICROALBUMINURIA AS A PREDICTOR OF CLINICAL DIABETIC NEPHROPATHY
    MOGENSEN, CE
    STEFFES, MW
    MYERS, BD
    VIBERTI, G
    FRIEDMAN, EA
    DECKERT, T
    MADIAS, NE
    ANDERSEN, OO
    MAUER, SM
    OSTERBY, R
    PARVING, HH
    [J]. KIDNEY INTERNATIONAL, 1987, 31 (02) : 673 - 689
  • [7] MOGENSEN CE, 1985, HYPERTENSION, V7, P64
  • [9] SULLIVAN K, 1991, DIABETES S1, V40, pA439
  • [10] EVALUATION OF MICRO-BUMINTEST REAGENT TABLETS FOR SCREENING OF MICROALBUMINURIA
    TAI, J
    TZE, WJ
    [J]. DIABETES RESEARCH AND CLINICAL PRACTICE, 1990, 9 (02) : 137 - 142