AMPEROMETRIC NEEDLE-TYPE GLUCOSE SENSOR BASED ON A MODIFIED PLATINUM-ELECTRODE WITH DIMINISHED RESPONSE TO INTERFERING MATERIALS

被引:43
作者
CHEN, CY
GOTOH, M
MAKINO, H
SU, YC
TAMIYA, E
KARUBE, I
机构
[1] UNIV TOKYO,ADV SCI & TECHNOL RES CTR,MEGURO KU,TOKYO 113,JAPAN
[2] NOK CORP,FUJISAWA,KANAGAWA,JAPAN
[3] CHIBA UNIV,SCH MED,DEPT INTERNAL MED 2,CHIBA,JAPAN
关键词
AMPEROMETRY; BIOSENSORS; ENZYMATIC METHODS; BLOOD; GLUCOSE; SERUM;
D O I
10.1016/0003-2670(92)85148-Y
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A needle-type glucose sensor that can be used to determine glucose in serum and whole blood samples was developed. Platinum wire was used as the working electrode and a disposable hypodermic stainless-steel needle electroplated with platinum was used as the counter and reference electrode. A method involving both photocross-linking of PVA-SbQ and cross-linking with glutaraldehyde was used to immobilize the enzyme [PVA-SbQ is a poly(vinyl alcohol) bearing stilbazolium groups]. Nafion and cellulose triacetate membranes were used to prevent inaccuracy from interfering materials and to increase the dynamic range of the sensor, respectively. The response, reproducibility and long-term stability of the sensor and the effects of temperature, pH and metal ions on the response were investigated. Owing to the effective method for enzyme immobilization, the large surface area of the counter electrode and the relative inactivity of the counter electrode to chemical reactions, the sensor showed good response, stability and reproducibility. The sensor did not respond to ascorbate and urate at the concentrations normally found in blood. Data obtained from the sensor for glucose in serum and whole blood samples showed a good correlation (r > 0.95) with a clinical laboratory automated analyser.
引用
收藏
页码:5 / 14
页数:10
相关论文
共 35 条
[1]   DETERMINATION OF GLUCOSE IN FERMENTATION PROCESSES BY MEANS OF AN ONLINE COUPLED FLOW-INJECTION SYSTEM USING ENZYME SENSORS BASED ON CHEMICALLY MODIFIED ELECTRODES [J].
APPELQVIST, R ;
HANSEN, EH .
ANALYTICA CHIMICA ACTA, 1990, 235 (02) :265-271
[2]   PULSED AMPEROMETRIC DETECTION OF GLUCOSE IN BIOLOGICAL-FLUIDS AT A SURFACE-MODIFIED GOLD ELECTRODE [J].
BINDRA, DS ;
WILSON, GS .
ANALYTICAL CHEMISTRY, 1989, 61 (22) :2566-2570
[3]   NEEDLE ENZYME ELECTRODES FOR BIOLOGICAL STUDIES [J].
CHURCHOUSE, SJ ;
BATTERSBY, CM ;
MULLEN, WH ;
VADGAMA, PM .
BIOSENSORS, 1986, 2 (06) :325-342
[4]   POTENTIALLY-IMPLANTABLE, FERROCENE-MEDIATED GLUCOSE SENSOR [J].
CLAREMONT, DJ ;
PENTON, C ;
PICKUP, JC .
JOURNAL OF BIOMEDICAL ENGINEERING, 1986, 8 (03) :272-274
[5]   IMPLANTED ELECTROENZYMATIC GLUCOSE SENSORS [J].
CLARK, LC ;
DUGGAN, CA .
DIABETES CARE, 1982, 5 (03) :174-180
[6]   IMMOBILIZATION OF BIOCATALYSTS WITH BOMBYX-MORI SILK FIBROIN BY SEVERAL KINDS OF PHYSICAL TREATMENT AND ITS APPLICATION TO GLUCOSE SENSORS [J].
DEMURA, M ;
ASAKURA, T ;
KUROO, T .
BIOSENSORS, 1989, 4 (06) :361-372
[7]   IMMOBILIZATION OF GLUCOSE-OXIDASE IN A POLY [VINYL ALCOHOL] MATRIX ON PLATINIZED GRAPHITE-ELECTRODES BY GAMMA-IRRADIATION [J].
GALIATSATOS, C ;
IKARIYAMA, Y ;
MARK, JE ;
HEINEMAN, WR .
BIOSENSORS & BIOELECTRONICS, 1990, 5 (01) :47-61
[8]   WHOLE-BLOOD GLUCOSE ENZYME ELECTRODE [J].
GENSHAW, MA ;
JONES, JE .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1989, 136 (02) :414-417
[9]   CROSS-LINKED REDOX GELS CONTAINING GLUCOSE-OXIDASE FOR AMPEROMETRIC BIOSENSOR APPLICATIONS [J].
GREGG, BA ;
HELLER, A .
ANALYTICAL CHEMISTRY, 1990, 62 (03) :258-263
[10]   FIBEROPTIC GLUCOSE SENSOR WITH ELECTROCHEMICAL GENERATION OF INDICATOR REAGENT [J].
GUNASINGHAM, H ;
TAN, CH ;
SEOW, JKL .
ANALYTICAL CHEMISTRY, 1990, 62 (07) :755-759