Electron beam characteristics at extended source-to-surface distances for irregular cut-outs

被引:10
作者
Arunkumar, T. [1 ]
Supe, Sanjay S. [1 ]
Ravikumar, M. [1 ]
Sathiyan, S. [1 ]
Ganesh, M. [1 ]
机构
[1] Kidwai Mem Inst Oncol, Dept Radiat Phys, Hosur Rd, Bangalore 560029, Karnataka, India
关键词
Cerrobend cutout; electron beam; flatness; isodose; uniformity index;
D O I
10.4103/0971-6203.71763
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Electron beam therapy is widely used in the management of cancers. The rapid dose fall-off and the short range of an electron beam enable the treatment of lesions close to the surface, while sparing the underlying tissues. In an extended source-to-surface (SSD) treatment with irregular field sizes defined by cerrobend cutouts, underdosage of the lateral tissue may occur due to reduced beam flatness and uniformity. To study the changes in the beam characteristics, the depth dose, beam profile, and isodose distributions were measured at different SSDs for regular 10 10 cm(2) and 15 15 cm(2) cone, and for irregular cutouts of field size 6.5 9 cm(2) and 11.5 15 cm(2) for beam energies ranging from 6 to 20 MeV. The PDD, beam flatness, symmetry and uniformity index were compared. For lower energy (6 MeV), there was no change in the depth of maximum dose (R100) as SSD increased, but for higher energy (20 MeV), the R100 depth increased from 2 cm to 3 cm as SSD increased. This shows that as SSD increases there is an increase in the depth of the maximum dose for higher energy beams. There is a 7 mm shift in the R100 depth when compared with regular and irregular field sizes. The symmetry was found to be within limits for all the field sizes as the treatment distance extended as per International Electro technical Commision (IEC) protocol. There was a loss of beam flatness for irregular fields and it was more pronounced for lower energies as compared with higher energies, so that the clinically useful isodose level (80% and 90%) width decreases with increase in SSD. This suggests that target coverage at extended SSD with irregular cut-outs may be inadequate unless relatively large fields are used.
引用
收藏
页码:207 / 214
页数:8
相关论文
共 8 条
[1]   ELECTRON-BEAM CHARACTERISTICS AT EXTENDED TREATMENT DISTANCES [J].
DAS, IJ ;
MCGEE, KP ;
CHENG, CW .
MEDICAL PHYSICS, 1995, 22 (10) :1667-1674
[2]   OPTIMUM FIELD SIZE AND CHOICE OF ISODOSE LINES IN ELECTRON-BEAM TREATMENT [J].
DAS, IJ ;
CHENG, CW ;
HEALEY, GA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 31 (01) :157-163
[3]  
Griffin J., 2006, RAD DOSIMETRY IRREGU
[4]  
ICRU, 1984, 35 ICRU
[5]  
International Electro technical Commission, 1984, 351 INT EL TECHN COM
[6]   CLINICAL ELECTRON-BEAM DOSIMETRY - REPORT OF AAPM RADIATION-THERAPY COMMITTEE TASK GROUP NO-25 [J].
KHAN, FM ;
DOPPKE, KP ;
HOGSTROM, KR ;
KUTCHER, GJ ;
NATH, R ;
PRASAD, SC ;
PURDY, JA ;
ROZENFELD, M ;
WERNER, BL .
MEDICAL PHYSICS, 1991, 18 (01) :73-109
[7]   DOSE DISTRIBUTION CONSIDERATIONS OF MEDIUM-ENERGY ELECTRON-BEAMS AT EXTENDED SOURCE-TO-SURFACE DISTANCE [J].
SHAW, CB ;
AYYANGAR, KM ;
PAWLICKI, T ;
KORB, LJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 32 (01) :159-164
[8]   Characteristics of bremsstrahlung in electron beams [J].
Zhu, TC ;
Das, IJ ;
Bjärngard, BE .
MEDICAL PHYSICS, 2001, 28 (07) :1352-1358