APPLICATION OF THE MBWC 2-DIMENSIONAL POSITION READOUT TECHNIQUE TO A MULTIWIRE PROPORTIONAL COUNTER

被引:24
|
作者
MIZOGAWA, T
SATO, M
AWAYA, Y
机构
[1] INST PHYS & CHEM RES,WAKO,SAITAMA 35101,JAPAN
[2] NAGAOKA UNIV TECHNOL,NAGAOKA,NIIGATA 940,JAPAN
来源
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT | 1995年 / 366卷 / 01期
关键词
D O I
10.1016/0168-9002(95)00682-6
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A two-dimensional position readout technique named ''modified backgammon method with weighted coupling capacitors (MBWC)'' has been applied to a multiwire proportional counter, and its image linearity and position resolution have been investigated. One of two types of the technique, the three-output MBWC method, has shown several kinds of image deformations. The sources of the deformation and the way to avoid them are clarified. On the other hand, an essentially linear image has been readily obtained by adopting the other type (four-output MBWC method). The overall spatial resolution parallel to the wire direction and the spatial resolution of the position encoding cathode for the direction perpendicular to the wire have been investigated with Fe K X-rays. The best values obtained are 130 and 110 mu m (FWHM), respectively, which correspond to similar to 600 x 600 pixels.
引用
收藏
页码:129 / 136
页数:8
相关论文
共 50 条
  • [41] 2-DIMENSIONAL READOUT OF A WIRE SPARK PLANE
    GUSTAVSSON, K
    HAGBERG, E
    KULLANDER, S
    MOBERG, HO
    EKELOF, T
    NUCLEAR INSTRUMENTS & METHODS, 1974, 117 (01): : 171 - 175
  • [42] SIMPLE POSITION-SENSITIVE PARALLEL-PLATE AVALANCHE DETECTOR WITH 2-DIMENSIONAL READOUT
    JARED, RC
    GLASSEL, P
    HUNTER, JB
    MORETTO, LG
    NUCLEAR INSTRUMENTS & METHODS, 1978, 150 (03): : 597 - 598
  • [43] Low pressure two dimensional position sensitive multiwire proportional chamber for the radioactive ion beam experiment
    Zhang, Bao-Guo
    Wen, Wan-Xin
    Wei, Zhi-Yong
    Zhang, Jin-Xia
    Feng, Ying
    Tan, Ji-Lian
    Jin, Gen-Ming
    Xiao, Zhi-Gang
    Wang, Hong-Wei
    Wu, He-Yu
    Li, Zu-Yu
    Wang, Su-Fang
    Hu, Rong-Jiang
    Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 2004, 38 (05): : 459 - 462
  • [44] HIGH-RESOLUTION READOUT OF MULTIWIRE PROPORTIONAL-COUNTERS USING CATHODE-COUPLED DELAY-LINE TECHNIQUE
    LACY, JL
    LINDSEY, RS
    NUCLEAR INSTRUMENTS & METHODS, 1974, 119 (03): : 483 - 498
  • [45] A MICROSTRIP GAS AVALANCHE CHAMBER WITH 2-DIMENSIONAL READOUT
    ANGELINI, F
    BELLAZZINI, R
    BREZ, A
    MASSAI, MM
    SPANDRE, G
    TORQUATI, MR
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1989, 283 (03): : 755 - 761
  • [46] Application of position-sensitive proportional counter to X-ray spectroscopy
    Hasegawa, T
    Fukumori, T
    Kaminishi, K
    Matsuda, T
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 2-LETTERS & EXPRESS LETTERS, 1996, 35 (9B): : L1225 - L1226
  • [47] A SINGLE-WIRE PROPORTIONAL COUNTER WITH DELAY-LINE POSITION READOUT FOR THE FOCAL PLANE OF A MAGNETIC SPECTROGRAPH
    KOIDE, K
    NUEVO, AB
    TAKAI, H
    MARECHAL, B
    DIETZSCH, O
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1983, 215 (1-2): : 177 - 182
  • [48] NEW STRUCTURE OF 2-DIMENSIONAL POSITION-SENSITIVE SEMICONDUCTOR DETECTOR AND APPLICATION
    YAMAMOTO, K
    YAMAGUCHI, S
    TERADA, Y
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1985, 32 (01) : 438 - 442
  • [49] APPLICATION OF THE 2-DIMENSIONAL TECHNIQUE TO THE CROSSING PAPER ELECTROPHORESIS OF IMMUNOLOGICAL REACTIONS
    NAKAMURA, S
    TAKEO, K
    KATUNO, A
    TOMINAGA, S
    CLINICA CHIMICA ACTA, 1959, 4 (06) : 893 - 900