COMPUTER-SIMULATION OF THE QUADRUPOLE MASS FILTER

被引:4
作者
TAKEBE, M
KUMASHIRO, S
机构
[1] Software Development Center, Shimadzu Corporation, Nakagyo-ku, Kyoto, 604
关键词
D O I
10.1016/0168-9002(95)00059-3
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A computer program, called MMFCODE, has been developed to analyze numerically the dynamics of ion beams in a Quadrupole Mass Filter (QMF). The motion of the ions can be known by solving the three-dimensional equation of motion. A harmonic expansion of the electric field is introduced to get a reasonable computing time. For the incident beam, more than 10(4) ions are generated by a Monte Carlo method. After the mass scan line was determined with the condition that Delta M = const. at two different mass numbers (M), the mass spectrum and the acceptance can be simulated. By using MMFCODE, the simulation of a real dimension can be done without neglecting the fringing field or using the model field. In this paper, a model of a QMF with an RF-only pre-filter, called the delayed DC ramp [W. Brubaker, U.S. patent 3,371,204 (1968)], was analyzed by using MMFCODE. By our simulation, the transmittance was oscillated with respect to the pre-filter length in the case of a low mass number as reported by C. Trajber et al. [Rapid Commun. Mass Spectrom. 6 (1992) 459], but was not oscillated in the case of a high mass number. This is probably because the distribution of the axial velocities is narrow in the case of a low mass number but broad in the case of a high mass number. This discrepancy is caused by the influence of the fringing field at the pre-filter entrance.
引用
收藏
页码:411 / 415
页数:5
相关论文
共 8 条
[1]  
BRUBAKER W, 1968, Patent No. 3371204
[2]   PERFORMANCE OF QUADRUPOLE MASS FILTER WITH SEPARATED RF AND DC FRINGING FIELDS [J].
DAWSON, PH .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1977, 25 (04) :375-392
[3]   OPERATING PARAMETERS OF A QUADRUPOLE IN A GROUNDED CYLINDRICAL HOUSING [J].
DENISON, DR .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY, 1971, 8 (01) :266-&
[4]  
HAIRER E, 1987, SOLVING ORDINARY DIF, V1, P180
[5]   INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY WITH AN ENLARGED SAMPLING ORIFICE AND OFFSET ION LENS .1. ION TRAJECTORIES AND DETECTOR PERFORMANCE [J].
HU, K ;
CLEMONS, PS ;
HOUK, RS .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1993, 4 (01) :16-27
[6]   FUNDAMENTAL-STUDY FOR HIGH-ACCURACY CALCULATION OF 3-D ELECTROMAGNETIC-FIELD [J].
KUNO, Y ;
YAGI, A ;
MORISHIMA, T ;
UCHIKAWA, Y .
IEEE TRANSACTIONS ON MAGNETICS, 1988, 24 (01) :295-298
[7]  
MAEDA R, 1992, NONLINEAR PHENOMENA, P533
[8]   A METHOD FOR UNIFORM OPTIMIZATION OF QUADRUPOLE PREFILTERS [J].
TRAJBER, C ;
SIMON, M ;
BOHATKA, S .