DESIGN AND OPERATION OF A UHV DOUBLE-BOUNCE HARMONIC REJECTION MIRROR FOR XAFS STUDIES

被引:4
作者
LAMBLE, GM
机构
关键词
D O I
10.1063/1.1145928
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A simple but effective harmonic rejection system has been installed on beamline X-11B at the NSLS. The system includes a parallel pair of nickel-coated float glass mirrors. This geometry allows an output beam whose direction is independent of the incidence angle on the mirrors. They are housed within a UHV chamber and situated 2.0 m upstream of the sample position and 8 m downstream of a double crystal monochromator. The angle of the mirror pair is adjusted by a single stepper motor. In its current configuration the double-bounce system works for energies between about 2 and 6 keV. For harmonic rejection of energies higher than this, the whole device may be repositioned to permit a single bounce of the x-ray beam from one of the mirrors. The features and operation of this simple device are discussed. © 1995 American Institute of Physics.
引用
收藏
页码:1422 / 1424
页数:3
相关论文
共 5 条
[1]   PERFECT-CRYSTAL MONOCHROMATORS FOR SYNCHROTRON X-RADIATION [J].
BONSE, U ;
MATERLIK, G ;
SCHRODER, W .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1976, 9 (JUN1) :223-230
[2]  
GOULON J, 1984, EXAFS NEAR EDGE STRU, V3, P449
[3]   SEPARATED FUNCTION FOCUSING MONOCHROMATOR SYSTEM FOR SYNCHROTRON RADIATION [J].
HASTINGS, JB ;
KINCAID, BM ;
EISENBERGER, P .
NUCLEAR INSTRUMENTS & METHODS, 1978, 152 (01) :167-171
[4]  
LYTLE FW, 1984, NUCL INSTRUM METH A, V256, P542
[5]   2-MIRROR DEVICE FOR HARMONIC REJECTION [J].
SAINCTAVIT, P ;
PETIAU, J ;
MANCEAU, A ;
RIVALLANT, R ;
BELAKHOVSKY, M ;
RENAUD, G .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (07) :2027-2029