INITIAL CLINICAL-EXPERIENCE WITH COMPUTER-CONTROLLED CONFORMAL RADIOTHERAPY OF THE PROSTATE USING A 50-MEV MEDICAL MICROTRON

被引:20
作者
MAGERAS, GS [1 ]
FUKS, Z [1 ]
OBRIEN, J [1 ]
BREWSTER, LJ [1 ]
BURMAN, C [1 ]
CHUI, CS [1 ]
LEIBEL, SA [1 ]
LING, CC [1 ]
MASTERSON, ME [1 ]
MOHAN, R [1 ]
KUTCHER, GJ [1 ]
机构
[1] MEM SLOAN KETTERING CANC CTR,DEPT RADIAT ONCOL,NEW YORK,NY 10021
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1994年 / 30卷 / 04期
关键词
CONFORMAL RADIOTHERAPY; COMPUTER-CONTROLLED THERAPY; MULTILEAF COLLIMATOR; PROSTATE;
D O I
10.1016/0360-3016(94)90374-3
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: We have described previously a model far delivering computer-controlled radiation treatments. We report here on the implementation and first year's clinical experience with such treatments using a 50 MeV medical microtron. Methods and Materials: The microtron is equipped with a multileaf collimator and is capable of setting up and treating a sequence of fixed fields called segments, under computer control. An external computer derives machine parameters for the segments from a three-dimensional treatment planning system, transfers them to the microtron control computer, checks the machine settings before allowing dose delivery to begin, and records the treatment. We describe the patient treatment methodology, portal film acquisition, electronic portal imaging, and quality assurance. Results: Patient treatments began in July 1992, comprising six-segment conformal treatments of the prostate. Using the recorded treatment data, the system performance has been examined and compared to other treatment machines. The average treatment time is 10 min, of which 4 min is for computer-controlled setup and irradiation; the remaining time is for patient positioning and checking of clearances. Long-term reproducibiIity of computer-controlled setup of the gantry and multileaf position is better than 0.5 degrees and 1 mm, respectively. Termination due to a machine fault has occurred in 5.5% of treatments, improving to 2.5% in recent months. Conclusion: Our initial experience indicates that computer-controlled segmental therapy can be performed reliably on a routine basis. Treatment times with the microtron are significantly shorter than with conventional linacs, and setup accuracy is consistent with that needed for conformal therapy. We believe that treatment times can be further improved through software upgrades and integration of electronic portal imaging.
引用
收藏
页码:971 / 978
页数:8
相关论文
共 15 条
  • [1] AUTOMATIC-GENERATION OF BEAM APERTURES
    BREWSTER, L
    MAGERAS, GS
    MOHAN, R
    [J]. MEDICAL PHYSICS, 1993, 20 (05) : 1337 - 1342
  • [2] Brewster Linda, 1992, International Journal of Radiation Oncology Biology Physics, V24, P157
  • [3] A COMPUTER-CONTROLLED RADIATION-THERAPY MACHINE FOR PELVIC AND PARA-AORTIC NODAL AREAS
    CHIN, LM
    KIJEWSKI, PK
    SVENSSON, GK
    CHAFFEY, JT
    LEVENE, MB
    BJARNGARD, BE
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1981, 7 (01): : 61 - 70
  • [4] CHUNGBIN A, 1979, 3RD P INT COMP SOFTW, P270
  • [5] JENNINGS WA, 1985, PROGR MED RAD PHYSIC, V2, P1
  • [6] Leibel, 1992, Semin Radiat Oncol, V2, P274, DOI 10.1016/1053-4296(92)90025-G
  • [7] IMPROVED DOSE DISTRIBUTIONS FOR 3D CONFORMAL BOOST TREATMENTS IN CARCINOMA OF THE NASOPHARYNX
    LEIBEL, SA
    KUTCHER, GJ
    HARRISON, LB
    FASS, DE
    BURMAN, CM
    HUNT, MA
    MOHAN, R
    BREWSTER, LJ
    LING, CC
    FUKS, ZY
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 20 (04): : 823 - 833
  • [8] LING CC, 1993, RADIOTHER ONCOL, V29, P129
  • [9] THE USE OF A MULTI-LEAF COLLIMATOR FOR CONFORMAL RADIOTHERAPY OF CARCINOMAS OF THE PROSTATE AND NASOPHARYNX
    LOSASSO, T
    CHUI, CS
    KUTCHER, GJ
    LEIBEL, SA
    FUKS, Z
    LING, CC
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 25 (02): : 161 - 170
  • [10] A MODEL FOR COMPUTER-CONTROLLED DELIVERY OF 3-D CONFORMAL TREATMENTS
    MAGERAS, GS
    PODMANICZKY, KC
    MOHAN, R