RHODIUM-IRON RESISTANCE THERMOMETER WITH FUSED-SILICA COIL FRAME

被引:5
作者
TAMURA, O
SAKURAI, H
机构
[1] National Research Laboratory of Metrology, Tsukuba, Ibaraki, 305, 1-4
关键词
THERMOMETRY; RESISTANCE THERMOMETER; RHODIUM IRON;
D O I
10.1016/0011-2275(91)90019-S
中图分类号
O414.1 [热力学];
学科分类号
摘要
The dependences of the thermometric properties of a rhodium - iron (Rh - Fe) (mole fraction 0.5%) wire resistance thermometer on the annealing temperature between 700 and 900-degrees-C are investigated using fused-silica coil frames. The thermometers annealed at or above 800-degrees-C have similar temperature - resistance characteristics. These thermometers can be calibrated with an accuracy of 0.5 mK over the range 4.2 - 25 K using a reference temperature - resistance function and three calibration points. Rh - Fe wire thermometers from different origins can be calibrated with an accuracy of the order of 10 mK in the same way. The temperature rise of the present thermometers due to heating by the measuring current is in a reasonable range for thermometry with a millikelvin accuracy.
引用
收藏
页码:869 / 873
页数:5
相关论文
共 6 条
[1]  
Meyers CH, 1932, BUR STAND J RES, V9, P807
[2]  
Rusby R. L., 1975, Temperature Measurement, 1975, P125
[3]  
Rusby R. L., 1972, TEMP ITS MEAS, V4, P865
[4]  
Rusby R.L., 1982, TEMP ITS MEAS, V5, P829
[5]  
SAWADA S, 1972, TEMP ITS MEAS, V4, P919
[6]   TEMPERATURE COEFFICIENTS OF RESISTIVITY OF RH-FE THIN-FILMS FOR CRYOGENIC THERMOMETER USE [J].
TAMURA, O ;
SAKURAI, H .
JAPANESE JOURNAL OF APPLIED PHYSICS PART 2-LETTERS, 1987, 26 (06) :L947-L950