DEVELOPMENT OF SEMIPORTABLE DC-SQUID MAGNETOMETER

被引:5
作者
KANDORI, A [1 ]
UEDA, M [1 ]
OGATA, H [1 ]
KADO, H [1 ]
机构
[1] ELECTROTECH LAB, TSUKUBA, IBARAKI 305, JAPAN
关键词
D O I
10.1109/77.403092
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A semi-portable DC-SQUID magnetometer with a liquid helium continuous how cryostat has been developed. The cryostat is 340 mm long and 50 mm in outer diameter. The pick-up coil is a co-axial, first order, differential type, 20 mm in diameter and 30 mm in baseline. The distance between the pick-up coil and the outer surface of the cryostat is 5 mm. It is easy to set this cryostat on the chest wall or on a human head. The cool down time is about 20 minutes, which is much shorter than that for a conventional storage Dewar. The white noise level, measured in a magnetically shielded room, is 8.7 fT/root Hz at 30 Hz. Noise at 1 Hz is about 30 fT/root Hz. Noise from temperature fluctuations (< 9 mK) and drift (4.22 K similar to 4.55 K) is negligible. The consumption rate of liquid helium is 10 l/h. In this paper, we show the measurement of a magnetocardiogram in a magnetically shielded room.
引用
收藏
页码:2474 / 2477
页数:4
相关论文
共 9 条
  • [1] INTEGRATED DC SQUID MAGNETOMETER WITH HIGH DV/DB
    DRUNG, D
    CANTOR, R
    PETERS, M
    RYHANEN, T
    KOCH, H
    [J]. IEEE TRANSACTIONS ON MAGNETICS, 1991, 27 (02) : 3001 - 3004
  • [2] LOW-NOISE HIGH-SPEED DC SUPERCONDUCTING QUANTUM INTERFERENCE DEVICE MAGNETOMETER WITH SIMPLIFIED FEEDBACK ELECTRONICS
    DRUNG, D
    CANTOR, R
    PETERS, M
    SCHEER, HJ
    KOCH, H
    [J]. APPLIED PHYSICS LETTERS, 1990, 57 (04) : 406 - 408
  • [3] VOLTAGE VERSUS FLUX RELATION OF DC SUPERCONDUCTING QUANTUM INTERFERENCE DEVICE USING 3 JOSEPHSON-JUNCTIONS
    ENPUKU, K
    DOI, H
    [J]. JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS BRIEF COMMUNICATIONS & REVIEW PAPERS, 1994, 33 (4A): : 1856 - 1862
  • [4] KANDORI A, UNPUB CONSIDERATIONS
  • [5] PRACTICAL LOW-NOISE INTEGRATED DC SUPERCONDUCTING QUANTUM INTERFERENCE DEVICE MAGNETOMETER WITH ADDITIONAL POSITIVE FEEDBACK
    RYHANEN, T
    CANTOR, R
    DRUNG, D
    KOCH, H
    [J]. APPLIED PHYSICS LETTERS, 1991, 59 (02) : 228 - 230
  • [6] DC-SQUID ELECTRONICS BASED ON ADAPTIVE POSITIVE FEEDBACK - EXPERIMENTS
    SEPPA, H
    AHONEN, A
    KNUUTILA, J
    SIMOLA, J
    VILKMAN, V
    [J]. IEEE TRANSACTIONS ON MAGNETICS, 1991, 27 (02) : 2488 - 2490
  • [7] TAKADA Y, 1994, P SOC PHOTO-OPT INS, V2160, P195, DOI 10.1117/12.181004
  • [8] PROXIMITY EFFECTS IN SUPERCONDUCTING WIRE BONDING
    Tsukada, K.
    Yamasaki, S.
    Mizutani, N.
    Uehara, G.
    Kado, H.
    Akimoto, H.
    Ogashiwa, T.
    [J]. IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1993, 3 (01) : 2965 - 2967
  • [9] DESIGN AND FABRICATION OF MULTICHANNEL DC SQUIDS FOR BIOMAGNETIC APPLICATIONS
    Yamasaki, S.
    Morooka, T.
    Matsuda, N.
    Kawai, J.
    Mizutani, N.
    Tsukada, K.
    Uehara, G.
    Kado, H.
    [J]. IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1993, 3 (01) : 1887 - 1889