PRECISION AND ACCURACY OF IN-VIVO BONE-MINERAL MEASUREMENT IN RATS USING DUAL-ENERGY X-RAY ABSORPTIOMETRY

被引:43
|
作者
ROZENBERG, S
VANDROMME, J
NEVE, J
AGUILERA, A
MUREGANCURO, A
PERETZ, A
KINTHAERT, J
HAM, H
机构
[1] Interdisciplinary Group on Osteoporosis, Free Universities of Brussels, Hôpital Saint-Pierre, Brussels, B-1000
关键词
ACCURACY; BONE MINERAL CONTENT; DXA; OSTEOPOROSIS; PRECISION;
D O I
10.1007/BF01623658
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The aim of this study was to evaluate the precision and accuracy of dual-energy X-ray absorptiometry (DXA) for measuring bone mineral content at different sites of the skeleton in rats. In vitro the reproducibility error was very small (< 1%), but in vivo the intra-observer variability ranged from 0.9% to 6.0%. Several factors have been shown to affect in vivo reproducibility: the reproducibility was better when the results were expressed as bone mineral density (BMD) rather than bone mineral content (BMC), intra-observer variability was better than the inter-observer variability, and a higher error was observed for the tibia compared with that for vertebrae and femur. The accuracy of measurement at the femur and tibia was assessed by comparing the values with ash weight and with biochemically determined calcium content. The correlation coefficients (R) between the in vitro BMC and the dry weight or the calcium content were higher than 0.99 for both the femur and the tibia. SEE ranged between 0.0 g (ash weight) and 2.0 mg (Ca content). Using in vitro BMC, ash weight could be estimated with an accuracy error close to 0 and calcium content with an error ranging between 0.82% and 6.80%. The R values obtained between the in vivo and in vitro BMC were 0.98 and 0.97 respectively for femur and tibia, with SEE of 0.04 and 0.02 g respectively. In conclusion, the in vivo precision of the technique was found to be too low. To be of practical use it is important in the design of experimentation to try to reduce the measurement error. This can be achieved by performing measurements in the same position, by repeating measurements several times and by using the mean values of several BMD calculations performed by the same observer on each BMD measurement. Furthermore, better reproducibility can be obtained on the vertebra or the femur than on the tibia.
引用
收藏
页码:47 / 53
页数:7
相关论文
共 50 条
  • [21] BONE-MINERAL DENSITY IN HEALTHY DUTCH WOMEN - SPINE AND HIP MEASUREMENTS USING DUAL-ENERGY X-RAY ABSORPTIOMETRY
    ERDTSIECK, RJ
    POLS, HAP
    ALGRA, D
    KOOY, PPM
    BIRKENHAGER, JC
    NETHERLANDS JOURNAL OF MEDICINE, 1994, 45 (05) : 198 - 205
  • [22] Calculating In-vivo Short-term Precision Error of Dual-Energy X-ray Absorptiometry in Human and Animal: A Technical Report
    Subramaniam, S.
    Mohamad, N., V
    Chan, C. Y.
    Soelaiman, I. N.
    Chin, K. Y.
    MEDICINE AND HEALTH-KUALA LUMPUR, 2020, 15 (01): : 70 - 77
  • [23] Precision, accuracy, and reproducibility of dual X-ray absorptiometry measurements in mice in vivo
    Iida-Klein, A
    Lu, SS
    Yokoyama, K
    Dempster, DW
    Nieves, JW
    Lindsay, R
    JOURNAL OF CLINICAL DENSITOMETRY, 2003, 6 (01) : 25 - 33
  • [24] Dual-energy X-ray Absorptiometry
    Jain, Rajesh K.
    Vokes, Tamara
    JOURNAL OF CLINICAL DENSITOMETRY, 2017, 20 (03) : 291 - 303
  • [25] TOTAL-BODY BONE-MINERAL AND PELVIS BONE-MINERAL CONTENT AS PARAMETERS OF BONE MASS IN MEN - A DUAL-ENERGY X-RAY ABSORPTIOMETRY STUDY
    HERNANDEZ, ER
    REVILLA, M
    RICO, H
    ACTA ANATOMICA, 1991, 142 (03): : 227 - 230
  • [26] Dual-Energy X-Ray Absorptiometry for Measurement of Phalangeal Bone Mineral Density on a Slot-Scanning Digital Radiography System
    Dendere, Ronald
    Potgieter, J. Herman
    Steiner, Stefan
    Whiley, Sarah P.
    Douglas, Tania S.
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2015, 62 (12) : 2850 - 2859
  • [27] COMPARISON OF NUCLEAR-MAGNETIC-RESONANCE SPECTROSCOPY WITH DUAL-PHOTON ABSORPTIOMETRY AND DUAL-ENERGY X-RAY ABSORPTIOMETRY IN THE MEASUREMENT OF THORACIC VERTEBRAL BONE-MINERAL DENSITY - COMPRESSIVE FORCE VERSUS BONE-MINERAL
    MYERS, TJ
    BATTOCLETTI, JH
    MAHESH, M
    GULATI, M
    WILSON, CR
    PINTAR, F
    REINARTZ, J
    OSTEOPOROSIS INTERNATIONAL, 1994, 4 (03) : 129 - 137
  • [28] Validation of peripheral dual-energy X-ray absorptiometry for the measurement of bone mineral in intact and excised long bones of rats
    Nagy, TR
    Prince, CW
    Li, J
    JOURNAL OF BONE AND MINERAL RESEARCH, 2001, 16 (09) : 1682 - 1687
  • [29] DUAL-ENERGY X-RAY ABSORPTIOMETRY OF THE CALCANEUS - COMPARISON WITH VERTEBRAL DUAL-ENERGY X-RAY ABSORPTIOMETRY AND QUANTITATIVE COMPUTED-TOMOGRAPHY
    LAVALJEANTET, AM
    BERGOT, C
    WILLIAMS, M
    DAVIDSON, K
    LAVALJEANTET, M
    CALCIFIED TISSUE INTERNATIONAL, 1995, 56 (01) : 14 - 18
  • [30] Precision assessment and radiation safety for dual-energy X-ray absorptiometry
    Baim, S
    Wilson, CR
    Lewiecki, EM
    Luckey, MM
    Downs, RW
    Lentle, BC
    JOURNAL OF CLINICAL DENSITOMETRY, 2005, 8 (04) : 371 - 378