A CHERENKOV VIEWING DEVICE FOR USED-FUEL VERIFICATION

被引:7
作者
ATTAS, EM
CHEN, JD
YOUNG, GJ
机构
[1] Atomic Energy of Canada Limited, Whiteshell Nuclear Research Establishment, Pinawa
关键词
D O I
10.1016/0168-9002(90)90753-S
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A Cherenkov viewing device (CVD) has been developed to help verify declared inventories of used nuclear fuel stored in water bays. The device detects and amplifies the faint ultraviolet Cherenkov glow from the water surrounding the fuel, producing a real-time visible image on a phosphor screen. Quartz optics, a UV-pass filter and a microchannel-plate image-intensifier tube serve to form the image, which can be photographed or viewed directly through an eyepiece. Normal fuel bay lighting does not interfere with the Cherenkov light image. The CVD has been successfully used to detect anomalous PWR, BWR and CANDU (CANada Deuterium Uranium: registered trademark) fuel assemblies in the presence of normal-burnup assemblies stored in used-fuel bays. The latest version of the CVD, known as Mark IV, is being used by inspectors from the International Atomic Energy Agency for verification of light-water power-reactor fuel. Its design and operation are described, together with plans for further enhancements of the instrumentation.
引用
收藏
页码:88 / 93
页数:6
相关论文
共 7 条
[1]  
CHEN JD, 1989, RC259SSP7 AT EN CAN
[2]  
Csorba I, 1985, BLACKSBURG CONTINUIN
[3]  
Jelley J.V, 1958, CHERENKOV RAD ITS AP
[4]  
NAKAOKA A, 1986, 1986 P INT S NUCL MA, P106
[5]  
NICHOLSON N, 1981, LA8767MS LOS AL NAT
[6]  
NILSSON A, 1989, 30TH P ANN M I NUCL, P902
[7]  
NILSSON A, 1989, 30TH P ANN M I NUCL, P910