MEASUREMENT OF PATIENT ENTRANCE SURFACE DOSE-RATES FOR FLUOROSCOPIC X-RAY UNITS

被引:21
作者
MARTIN, CJ
机构
[1] Dept. of Biomed. Phys. and Bioeng., Aberdeen Royal Hospitals-NHS Trust
关键词
D O I
10.1088/0031-9155/40/5/008
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Measurements of patient entrance surface dose rate provide valuable data for interpreting results from dose-area product studies on fluoroscopic x-ray equipment. Methods for measurement of entrance surface dose rate with backscatter and incident dose rate without backscatter have been investigated. Entrance surface dose rate is measured with an ionization chamber in contact with a tissue-equivalent phantom. Backscattered radiation contributes 27-45% to the measurement and is affected by field size and chamber position. Incident dose rare measured using a copper phantom provides an alternative approach. Consistent relationships between thicknesses of Perspex and copper giving similar incident dose rates under automatic gain control have been established for different tube potentials with and without a grid. This allows measurements of incident dose rate made using copper to be linked to corresponding thicknesses of tissue-equivalent material. Since only a few millimetres of copper are required, contributions from backscatter can be minimized and transport of phantoms is simplified. Incident dose can be related to dose-area product and entrance surface dose derived using backscatter factors. Such measurements play a valuable role in interpreting patient dose data and recommending options to reduce patient dose.
引用
收藏
页码:823 / 834
页数:12
相关论文
共 22 条
[1]   DIFFERENCES IN REPORTED BACKSCATTER FACTORS FOR LOW-ENERGY X-RAYS - A LITERATURE STUDY [J].
CARLSSON, CA .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (04) :521-531
[2]   PHANTOM EQUIVALENT THICKNESSES IN DIAGNOSTIC-RADIOLOGY [J].
CARRIER, R ;
BLAIS, N .
PHYSICS IN MEDICINE AND BIOLOGY, 1987, 32 (09) :1151-1157
[3]  
CRANLEY K, 1991, IPSM64 REP
[4]   RADIATION-DOSE RECEIVED BY PATIENTS DURING BARIUM MEAL INVESTIGATIONS UNDER AUTOMATIC BRIGHTNESS CONTROL [J].
FAULKNER, K ;
BRAMALL, G .
BRITISH JOURNAL OF RADIOLOGY, 1985, 58 (685) :31-34
[5]   BACKSCATTER FACTORS FOR X-RAYS GENERATED AT VOLTAGES BETWEEN 10 AND 100 KV [J].
GROSSWENDT, B .
PHYSICS IN MEDICINE AND BIOLOGY, 1984, 29 (05) :579-591
[6]   DEPENDENCE OF THE PHOTON BACKSCATTER FACTOR FOR WATER ON SOURCE-TO-PHANTOM DISTANCE AND IRRADIATION FIELD SIZE [J].
GROSSWENDT, B .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (09) :1233-1245
[7]   BACKSCATTER FACTORS FOR DIAGNOSTIC-RADIOLOGY (1-4 MM AL HVL) [J].
HARRISON, RM .
PHYSICS IN MEDICINE AND BIOLOGY, 1982, 27 (12) :1465-1474
[8]   PATIENT RADIATION-DOSES FROM ENTEROCLYSIS EXAMINATIONS [J].
HART, D ;
HAGGETT, PJ ;
BOARDMAN, P ;
NOLAN, DJ ;
WALL, BF .
BRITISH JOURNAL OF RADIOLOGY, 1994, 67 (802) :997-1000
[9]   ESTIMATION OF EFFECTIVE DOSE FROM DOSE-AREA PRODUCT MEASUREMENTS FOR BARIUM MEALS AND BARIUM ENEMAS [J].
HART, D ;
WALL, BF .
BRITISH JOURNAL OF RADIOLOGY, 1994, 67 (797) :485-489
[10]  
HART D, 1994, NRPB R262 REP