The quality assurance of volumetric modulated arc therapy (VMAT) plans for early stage prostate cancer: a technical note

被引:4
作者
Elith, Craig A. [1 ,2 ]
Dempsey, Shane E. [2 ]
Cao, Fred [1 ]
Farshadi, Afrooz [1 ]
Warren-Forward, Helen M. [2 ]
机构
[1] British Columbia Canc Agcy, Fraser Valley Ctr, Surrey, BC, Canada
[2] Univ Newcastle, Sch Hlth Sci, Callaghan, NSW, Australia
来源
JOURNAL OF MEDICAL RADIATION SCIENCES | 2014年 / 61卷 / 04期
关键词
ArcCHECK; IMSure; quality assurance; VMAT;
D O I
10.1002/jmrs.78
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
As radiation therapy transitions from intensity modulated radiation therapy (IMRT) to volumetric modulated arc therapy (VMAT) it is important to consider the quality assurance (QA) of VMAT plans in light of what has previously been learned and developed in IMRT QA. This technical note assesses if IMRT based plan QA software, which has reduced the need in IMRT for phantom dose measurements on the linear accelerator, can be incorporated into VMAT QA processes. Twenty prostate cases were retrospectively planned using VMAT with one arc to deliver a prescription of 74 Gy in 37 fractions. A plan QA was performed using both IMSure (version 3.3), a software-based IMRT QA program, and ArcCHECK (version 6.2.3.5713), a phantom-based VMAT QA tool. Outcomes assessed included the time needed to perform the QA of both the IMSure and ArcCHECK QA methods, and agreement between planned dose and QA measured dose. On average per case, the ArcCHECK technique needed 31.5 min to perform the VMAT plan QA, while IMSure required 3.5 min to perform the same QA. All 20 cases passed dosimetric QA using ArcCHECK. However, using IMSure, three cases failed dosimetric QA using the departments existing IMRT QA criteria. This research has demonstrated that the IMRT QA software IMSure may be incorporated into the QA of VMAT plans, however the criteria to assess the dosimetry of the VMAT plans may need to be different to that for IMRT cases. The implication of this research for radiation therapists is to be critically aware of the differences between the plan QA requirements and methods for IMRT and those required for VMAT.
引用
收藏
页码:261 / 266
页数:6
相关论文
共 18 条
[1]  
[Anonymous], 2007, WHY TREATMENT PLAN Q
[2]   Dosimetric validation of Acuros® XB with Monte Carlo methods for photon dose calculations [J].
Bush, K. ;
Gagne, I. M. ;
Zavgorodni, S. ;
Ansbacher, W. ;
Beckham, W. .
MEDICAL PHYSICS, 2011, 38 (04) :2208-2221
[3]   A Retrospective Planning Analysis Comparing Volumetric-Modulated Arc Therapy (VMAT) to Intensity-Modulated Radiation Therapy (IMRT) for Radiotherapy Treatment of Prostate Cancer [J].
Elith, Craig A. ;
Cao, Fred ;
Dempsey, Shane E. ;
Findlay, Naomi ;
Warren-Forward, Helen .
JOURNAL OF MEDICAL IMAGING AND RADIATION SCIENCES, 2013, 44 (02) :79-86
[4]   Comparing four volumetric modulated arc therapy beam arrangements for the treatment of early-stage prostate cancer [J].
Elith, Craig A. ;
Dempsey, Shane E. ;
Warren-Forward, Helen M. .
JOURNAL OF MEDICAL RADIATION SCIENCES, 2014, 61 (02) :91-101
[5]   A retrospective planning analysis comparing intensity modulated radiation therapy (IMRT) to volumetric modulated arc therapy (VMAT) using two optimization algorithms for the treatment of early-stage prostate cancer [J].
Elith, Craig A. ;
Dempsey, Shane E. ;
Warren-Forward, Helen M. .
JOURNAL OF MEDICAL RADIATION SCIENCES, 2013, 60 (03) :84-92
[6]   IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119 [J].
Ezzell, Gary A. ;
Burmeister, Jay W. ;
Dogan, Nesrin ;
LoSasso, Thomas J. ;
Mechalakos, James G. ;
Mihailidis, Dimitris ;
Molineu, Andrea ;
Palta, Jatinder R. ;
Ramsey, Chester R. ;
Salter, Bill J. ;
Shi, Jie ;
Xia, Ping ;
Yue, Ning J. ;
Xiao, Ying .
MEDICAL PHYSICS, 2009, 36 (11) :5359-5373
[7]  
Fan J, 2009, USE IMSURE MU VERIFI
[8]  
Fogliata A, 2011, PHYS MED BIOL, V38, P2208
[9]   Measurement comparison and Monte Carlo analysis for volumetric-modulated arc therapy (VMAT) delivery verification using the ArcCHECK dosimetry system [J].
Lin, Mu-Han ;
Koren, Sion ;
Veltchev, Iavor ;
Li, Jinsheng ;
Wang, Lu ;
Price, Robert A. ;
Ma, C. -M. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (02) :220-233
[10]   A technique for the quantitative evaluation of dose distributions [J].
Low, DA ;
Harms, WB ;
Mutic, S ;
Purdy, JA .
MEDICAL PHYSICS, 1998, 25 (05) :656-661