BACKGROUND REJECTION AND SIGNAL-TO-NOISE OPTIMIZATION IN CONFOCAL AND ALTERNATIVE FLUORESCENCE MICROSCOPES

被引:99
作者
SANDISON, DR [1 ]
WEBB, WW [1 ]
机构
[1] CORNELL UNIV, DEPT PHYS, ITHACA, NY 14853 USA
来源
APPLIED OPTICS | 1994年 / 33卷 / 04期
关键词
CONFOCAL MICROSCOPE; SIGNAL; BACKGROUND; NOISE; NIPKOW DISK; LINE ILLUMINATION; SLIT DETECTION; FLUORESCENCE; RESOLUTION VOLUME;
D O I
10.1364/AO.33.000603
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
In the confocal microscope, tightly focused illumination and spatially filtered detection are combined to reduce out-of-focus background and to produce high-quality images that display thin optical sections within thick fluorescent specimens. We define background as the detected light that originates outside a resolution volume and signal as the detected light that originates within the same volume. Background rejection is measured by the signal-to-background ratio (S/B) and is calculated for confocal, spinning-disk, line-illumination, slit-detection, and conventional fluorescence microscopes as a function of both the spatial filter size and the specimen thickness. Spatial filter sizes that reject background and optimize the signal-to-noise ratio (S/N) are calculated for each microscope. These calculations are normalized so that the time-averaged illumination at each point in the specimen is the same for each microscope. For thick specimens, we show that the S/B obtained with a confocal microscope can be more than 100 times greater than the S/B available with a conventional microscope, and we find that the optimal confocal S/N can be a factor of 10 greater than the S/N in the conventional microscope.
引用
收藏
页码:603 / 615
页数:13
相关论文
共 26 条
[1]  
[Anonymous], [No title captured], Patent No. [3013467 A, 3013467]
[2]   DEGREES OF FREEDOM, AND EIGENFUNCTIONS, FOR NOISY IMAGE [J].
BENDINELLI, M ;
CONSORTINI, A ;
RONCHI, L ;
FRIEDEN, BR .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1974, 64 (11) :1498-1502
[3]  
BORN M, 1980, PRINCIPLES OPTICS, P437
[4]  
Brakenhoff GJ, 1990, MICRO 90
[5]  
CARRINGTON WA, 1990, HDB BIOL CONFOCAL MI, pCH14
[6]   INFORMATION CAPACITY AND RESOLUTION IN AN OPTICAL-SYSTEM [J].
COX, IJ ;
SHEPPARD, CJR .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1986, 3 (08) :1152-1158
[7]   NEW REFLECTED-LIGHT MICROSCOPE FOR VIEWING UNSTAINED BRAIN AND GANGLION CELLS [J].
EGGER, MD ;
PETRAN, M .
SCIENCE, 1967, 157 (3786) :305-&
[8]  
GRADSHTEYN IS, 1980, TABLE INTEGRALS SERI, P682
[9]   DETERMINATION OF 3-DIMENSIONAL IMAGING PROPERTIES OF A LIGHT-MICROSCOPE SYSTEM - PARTIAL CONFOCAL BEHAVIOR IN EPIFLUORESCENCE MICROSCOPY [J].
HIRAOKA, Y ;
SEDAT, JW ;
AGARD, DA .
BIOPHYSICAL JOURNAL, 1990, 57 (02) :325-333
[10]   AN IMPROVED METHOD FOR COMPUTING A DISCRETE HANKEL TRANSFORM [J].
JOHNSON, HF .
COMPUTER PHYSICS COMMUNICATIONS, 1987, 43 (02) :181-202