MONTE-CARLO MODEL OF PHOTON BEAMS USED IN RADIATION-THERAPY

被引:98
作者
LOVELOCK, DMJ
CHUI, CS
MOHAN, R
机构
[1] Department of Medical Physics, Memorial Sloan-Kettering Cancer Center, New York
关键词
MONTE CARLO SIMULATION; RADIATION THERAPY MACHINE; DOSE CALCULATION;
D O I
10.1118/1.597620
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A generic Monte Carlo model of a photon therapy machine is described. The model, known as McRad, is based on EGS4 and has been in use since 1991. Its primary function has been the characterization of the incident photon fluence for use by dose calculation algorithms. The accuracy of McRad is examined by comparing the dose distributions in a water phantom generated using only the Monte Carlo data with measured dose distributions for two machines in our clinic; a 6 MV Varian Clinac 600C and the 15 MV beam from a Clinac 2100C. The Monte Carlo generated dose distributions are computed using a dose calculation algorithm based on the use of differential pencil beam kernels. It was found that the match to measured data could be improved if the model is tuned by adjusting the energy of the electron beam incident on the target. The beam profiles were found to be more sensitive indicators of the electron beam energy than the depth dose curves. Beyond the depths reached by contaminant electrons, the computed and measured depth dose curves agree to better than 1%. The comparison of beam profiles indicate that in regions up to within 1 cm of the field edge, the measured and computed doses generally agree to within 2%-3%.
引用
收藏
页码:1387 / 1394
页数:8
相关论文
共 15 条
  • [1] BIELAJEW AF, 1987, NUCL INSTRUM METH B, V18, P165
  • [2] BIELAJEW AF, 1989, PIRS0203 NAT RES COU
  • [3] MONTE-CARLO STUDY OF ACCELERATOR HEAD SCATTER
    CHANEY, EL
    CULLIP, TJ
    GABRIEL, TA
    [J]. MEDICAL PHYSICS, 1994, 21 (09) : 1383 - 1390
  • [4] ANGULAR-DISTRIBUTION OF BREMSSTRAHLUNG FROM 15-MEV ELECTRONS INCIDENT ON THICK TARGETS OF BE, AL, AND PB
    FADDEGON, BA
    ROSS, CK
    ROGERS, DWO
    [J]. MEDICAL PHYSICS, 1991, 18 (04) : 727 - 739
  • [5] ANALYSIS OF THE PHOTON-BEAM TREATMENT PLANNING DATA FOR A SCANNING BEAM MACHINE
    LOVELOCK, DMJ
    CHUI, CS
    KUTCHER, GJ
    MOHAN, R
    [J]. MEDICAL PHYSICS, 1994, 21 (12) : 1969 - 1977
  • [6] LOVELOCK DMJ, 1991, AAPM ANN GENERAL M
  • [7] CHARACTERISTICS OF THE HIGH-ENERGY PHOTON-BEAM OF A 25-MEV ACCELERATOR
    LUXTON, G
    ASTRAHAN, MA
    [J]. MEDICAL PHYSICS, 1988, 15 (01) : 82 - 87
  • [8] MACKIE TR, 1984, MED PHYS, V12, P188
  • [9] ENERGY AND ANGULAR-DISTRIBUTIONS OF PHOTONS FROM MEDICAL LINEAR ACCELERATORS
    MOHAN, R
    CHUI, C
    LIDOFSKY, L
    [J]. MEDICAL PHYSICS, 1985, 12 (05) : 592 - 597
  • [10] MOHAN R, 1992, ADV RAD ONCOLOGY PHY, P307