Direct and indirect loading of the Ilizarov external fixator: the effect on the interfragmentary movements and compressive loads

被引:15
作者
Gessmann, Jan [1 ]
Baecker, Hinnerk [1 ]
Jettkant, Birger [2 ]
Muhr, Gert [1 ]
Seybold, Dominik [1 ]
机构
[1] BG Univ Klinikum Bergmannsheil, Dept Surg & Traumatol, Burkle-De-La-Camp-Pl 1, D-44789 Bochum, Germany
[2] BG Univ Klinikum Bergmannsheil, Dept Surg Res, D-44789 Bochum, Germany
来源
STRATEGIES IN TRAUMA AND LIMB RECONSTRUCTION | 2011年 / 6卷 / 01期
关键词
Ilizarov; External fixator; Weight bearing; Indirect loading; Interfragmentary movement;
D O I
10.1007/s11751-011-0103-6
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The amount of weight bearing and the force transmission to the frame have an important influence on the results of treatment with an Ilizarov external fixator. The frame provides beneficial interfragmentary movements and compressive loads at the fracture site through elastic wires. Mobilisation can be achieved by applying a weight-bearing platform at the distal end of the fixator. The effect on the interfragmentary movements and the compressive loads in indirect and direct loading were analysed in this study using a composite tibia bone model. Displacement transducers were attached to measure the interfragmentary movements and to detect relative movements of the bone fragments and movements between the rings. The compressive loads in the osteotomy were measured with loading cells in the defect zone. The weight-bearing platform had a substantial effect on the biomechanical behaviour of the frame. It led to an indirect force transmission through the fixator with respect to the osteotomy, resulting in lower compressive loads, lower interfragmentary movements and higher mechanical stress on the frame.
引用
收藏
页码:27 / 31
页数:5
相关论文
共 19 条
[1]  
ARO HT, 1993, CLIN ORTHOP RELAT R, P8
[2]  
ARONSON J, 1992, CLIN ORTHOP RELAT R, P23
[3]   Stability of external circular fixation: a multi-variable biomechanical analysis [J].
Bronson, DG ;
Samchukov, ML ;
Birch, JG ;
Browne, RH ;
Ashman, RB .
CLINICAL BIOMECHANICS, 1998, 13 (06) :441-448
[4]   EFFECT OF DYNAMIZATION ON GAP HEALING OF DIAPHYSEAL FRACTURES UNDER EXTERNAL FIXATION [J].
CLAES, LE ;
WILKE, HJ ;
AUGAT, P ;
RUBENACKER, S ;
MARGEVICIUS, KJ .
CLINICAL BIOMECHANICS, 1995, 10 (05) :227-234
[5]   Ilizarov wire tensioning and holding methods: a biomechanical study [J].
Davidson, AW ;
Mullins, M ;
Goodier, D ;
Barry, M .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2003, 34 (02) :151-154
[6]  
Duda GN, 2000, UNFALLCHIRURG, V103, P839, DOI 10.1007/s001130050630
[7]   Interfragmentary movements in the early phase of healing in distraction and correction osteotomies stabilize with ring fixators [J].
Duda, GN ;
Sporrer, S ;
Sollmann, M ;
Hoffmann, JE ;
Kassi, JP ;
Khodadadyan, C ;
Raschke, M .
LANGENBECKS ARCHIVES OF SURGERY, 2003, 387 (11-12) :433-440
[8]  
FLEMING B, 1989, CLIN ORTHOP RELAT R, P95
[9]   THE INFLUENCE OF INDUCED MICROMOVEMENT UPON THE HEALING OF EXPERIMENTAL TIBIAL FRACTURES [J].
GOODSHIP, AE ;
KENWRIGHT, J .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1985, 67 (04) :650-655
[10]  
ILIZAROV GA, 1988, B HOSP JOINT DIS ORT, V48, P1