CLINICAL-RESULTS OF A 15-FILTER DIGITAL AUDITORY PROSTHESIS USING SELECTIVE AMPLIFICATION AND COMPRESSION

被引:6
作者
CHOUARD, CH [1 ]
GENIN, J [1 ]
MEYER, B [1 ]
机构
[1] CHU PARIS ST ANTOINE,RECH OTORHINOLARYNGOL LAB,PARIS,FRANCE
关键词
FREQUENCY RESPONSE; HEARING AID; DIGITAL SYSTEM;
D O I
10.1080/00016489.1992.11665410
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Classical hearing aid efficiency was compared with that of a 15-filter digital auditory prosthesis in 23 patients suffering from partial sensorineural hearing loss. This device contains a microprocessor 320C25 from TEXAS Instruments. The sound signal is analysed at a sampling rate of 1 6 kHz. With the use of a Fast Fourier Transform (FFT), it provides the system with 15 energy levels at frequencies ranging from 0 to 8 kHz. The energy of each frequency band is multiplied by a coefficient stored in the device memory and directly dependent on the patient's audiometric specificities. The 15 signals are then synthesized by inverse FFT and mixed in a loudspeaker. The prosthesis is a rechargable battery powered 4 x 10 x 15 cm box programmable through a PC. The clinical results obtained with this 15-filter digital prosthesis are discussed and compared with the performances of present classical hearing aids.
引用
收藏
页码:230 / 236
页数:7
相关论文
共 9 条
[1]  
BITJOKA L, 1990, CONTRIBUTION MISE PO
[2]  
Chouard C H, 1988, Ann Otolaryngol Chir Cervicofac, V105, P227
[3]  
CHOUARD CH, 1977, ANN OTO-LAR CHIR C-F, V94, P353
[4]  
GEGU D, 1984, THESIS U PARIS 12
[5]  
JUVANON JM, 1985, THESIS U PARIS SUD
[6]  
KATES J M, 1990, Journal of Rehabilitation Research and Development, V27, P255, DOI 10.1682/JRRD.1990.07.0255
[7]  
LEWITT H, 1990, ACTA OTOLARYNGOL STO, V469, P57
[8]  
RABINER LR, 1974, THEORY APPLICATION D
[9]  
VENTADOUR R, 1987, THESIS U PARIS 12