TRANSCUTANEOUS HYDROGEN CLEARANCE - A NEW NONINVASIVE TECHNIQUE FOR ASSESSING BLOOD-FLOW IN HUMAN SKIN

被引:3
|
作者
HARRISON, DK [1 ]
RAAD, RA [1 ]
NEWTON, D [1 ]
MCCOLLUM, PT [1 ]
机构
[1] UNIV DUNDEE,SCH BIOMED ENGN,DUNDEE DD1 4HN,SCOTLAND
关键词
D O I
10.1088/0967-3334/15/1/008
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A new technique for the non-invasive assessment of skin blood flow is described using a modified, commercially available transcutaneous oxygen electrode. Initially, experiments were carried out on normal volunteers at skin temperatures of 40 degrees C and 44 degrees C. Transcutaneous H-2 (tcpH(2)) clearances were measured after the subject had breathed a mixture of 2% H-2 in air until a saturation value in the skin had been reached, tcpH(2) clearance results were compared with laser Doppler flux values measured simultaneously on adjacent sites and at the same temperatures under control and reactive hyperaemic conditions. Good reproducibility for the tcpH(2) technique and significant correlation (p < 0.002) between it and laser Doppler flux values at the same temperatures was demonstrated. In further experiments, (I-125) 4-iodoantipyrine clearances (IAP) were carried out immediately prior to and at the same sites as tcpH(2) clearances in critically ischaemic lower limbs. The correlation between the methods and the differences between tcpH(2) values observed in normal and critically ischaemic skin were highly significant (p < 0.002).
引用
收藏
页码:89 / 100
页数:12
相关论文
共 50 条
  • [1] THE ISOTOPE CLEARANCE TECHNIQUE FOR MEASURING SKIN BLOOD-FLOW
    YOUNG, CMA
    HOPEWELL, JW
    BRITISH JOURNAL OF PLASTIC SURGERY, 1983, 36 (02): : 222 - 230
  • [2] LASER DOPPLER FLOWMETRY - A NEW TECHNIQUE FOR NONINVASIVE ASSESSMENT OF SKIN BLOOD-FLOW
    NILSSON, GE
    SALERUD, EG
    TENLAND, T
    OBERG, PA
    WAHLBERG, JE
    COSMETICS & TOILETRIES, 1984, 99 (03): : 97 - &
  • [3] HYDROGEN CLEARANCE - ASSESSMENT OF TECHNIQUE FOR MEASUREMENT OF SKIN-FLAP BLOOD-FLOW IN PIGS
    THOMSON, JG
    KERRIGAN, CL
    PLASTIC AND RECONSTRUCTIVE SURGERY, 1991, 88 (04) : 657 - 663
  • [4] USE OF THE HYDROGEN CLEARANCE TECHNIQUE FOR MEASUREMENTS OF PANCREATIC BLOOD-FLOW
    MACHENS, HG
    SENNINGER, N
    RUNKEL, N
    FRANK, G
    VONKUMMER, R
    HERFARTH, C
    JOURNAL OF SURGICAL RESEARCH, 1993, 55 (02) : 122 - 130
  • [5] RECENT ADVANCES IN NONINVASIVE ASSESSMENT OF HUMAN SKIN BLOOD-FLOW
    AGACHE, PG
    DUPOND, AS
    ACTA DERMATO-VENEREOLOGICA, 1994, : 47 - 51
  • [6] EVALUATION OF AN ISOTOPIC CLEARANCE TECHNIQUE FOR MEASUREMENT OF SKIN BLOOD-FLOW IN PIG
    MOUSTAFA, HF
    HOPEWELL, JW
    MICROVASCULAR RESEARCH, 1976, 11 (02) : 147 - 153
  • [7] SPINAL-CORD BLOOD-FLOW MEASURED BY A HYDROGEN CLEARANCE TECHNIQUE
    GRIFFITHS, IR
    ROWAN, JO
    CRAWFORD, RA
    JOURNAL OF THE NEUROLOGICAL SCIENCES, 1975, 26 (04) : 529 - 544
  • [8] NEW DOPPLER TECHNIQUE FOR ASSESSING UTEROPLACENTAL BLOOD-FLOW
    CAMPBELL, S
    GRIFFIN, DR
    PEARCE, JM
    DIAZRECASENS, J
    COHENOVERBEEK, TE
    WILLSON, K
    TEAGUE, MJ
    LANCET, 1983, 1 (8326): : 675 - 677
  • [9] ADVANTAGES AND DISADVANTAGES OF USING THE HYDROGEN CLEARANCE TECHNIQUE TO MEASURE PANCREATIC BLOOD-FLOW
    MACHENS, HG
    SENNINGER, N
    RUNKEL, N
    FRANK, G
    VONKUMMER, R
    HERFARTH, C
    EUROPEAN JOURNAL OF SURGERY, 1992, 158 (02) : 113 - 116
  • [10] ISOTOPE CLEARANCE TECHNIQUES IN MEASURING SKIN BLOOD-FLOW
    YOUNG, CMA
    HOPEWELL, JW
    BRITISH JOURNAL OF PLASTIC SURGERY, 1983, 36 (02): : 219 - 219