ASSESSMENT OF WHOLE-BODY COMPOSITION WITH DUAL-ENERGY X-RAY ABSORPTIOMETRY

被引:134
|
作者
SLOSMAN, DO
CASEZ, JP
PICHARD, C
ROCHAT, T
FERY, F
RIZZOLI, R
BONJOUR, JP
MORABIA, A
DONATH, A
机构
[1] HOP CANTONAL UNIV,CLIN EPIDEMIOL UNIT,CH-1211 GENEVA 14,SWITZERLAND
[2] HOP CANTONAL UNIV,DIV RESP GASTROENTEROL & NUTR,CH-1211 GENEVA 14,SWITZERLAND
[3] HOP CANTONAL UNIV,DIV RESP DIS,CH-1211 GENEVA 14,SWITZERLAND
[4] HOP CANTONAL UNIV,DIV CLIN PATHOPHYSIOL,CH-1211 GENEVA 14,SWITZERLAND
[5] UNIV BERN,MED POLYCLIN,CH-3000 BERN,SWITZERLAND
关键词
ACQUIRED IMMUNODEFICIENCY SYNDROME (AIDS); BONES; ABSORPTIOMETRY; FIBROSIS; CYSTIC; NUTRITION;
D O I
10.1148/radiology.185.2.1410379
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Dual-energy x-ray absorptiometry (DXA) allows noninvasive direct measurement of the three major components of body composition: lean body mass (LBM), fat body mass (FBM), and bone mineral body mass (BBM). To study the accuracy and short-term and long-term precision of body composition measurements, the authors measured body composition with DXA in 60 healthy young adults. Independent measurement of LBM (LBM(K-40)), obtained from the determination of the whole-body content of potassium-40 with a whole-body scintillation detector, was highly correlated with LBM determined with DXA (LBM(DXA)) (LBM(K-40) = 1.069 . LBM(DXA), R2 = .996). Assessment of body composition in 10 patients with acquired immunodeficiency syndrome (AIDS) and 10 patients with cystic fibrosis was performed with DXA. The AIDS patients showed a marked decrease in LBM, while in patients with cystic fibrosis LBM and BBM were decreased. DXA measurements of body composition appeared accurate and precise enough to be of clinical relevance in detecting specific alterations of body composition.
引用
收藏
页码:593 / 598
页数:6
相关论文
共 50 条
  • [1] Whole-body dual-energy X-ray absorptiometry
    Braillon, Pierre
    PEDIATRIC RADIOLOGY, 2009, 39 (11) : 1256 - 1257
  • [2] Whole-body dual-energy X-ray absorptiometry
    Pierre Braillon
    Pediatric Radiology, 2009, 39 : 1256 - 1257
  • [3] Body composition assessment by dual-energy X-ray absorptiometry (DXA)
    Andreoli, A.
    Scalzo, G.
    Masala, S.
    Tarantino, U.
    Guglielmi, G.
    RADIOLOGIA MEDICA, 2009, 114 (02): : 286 - 300
  • [4] Dual-energy X-ray absorptiometry and body composition
    Laskey, MA
    NUTRITION, 1996, 12 (01) : 45 - 51
  • [5] Dual-energy X-ray absorptiometry and body composition
    Plank, LD
    CURRENT OPINION IN CLINICAL NUTRITION AND METABOLIC CARE, 2005, 8 (03): : 305 - 309
  • [6] Pediatric software for whole-body dual-energy x-ray absorptiometry: Changes in bone and body composition estimates
    Ellis, KJ
    Shypailo, RJ
    FASEB JOURNAL, 2001, 15 (05): : A1003 - A1003
  • [7] In vivo validation of whole-body composition estimates from dual-energy X-ray absorptiometry.
    Prior, B
    Cureton, K
    Modlesky, C
    Evans, E
    Sloniger, M
    Saunders, M
    Lewis, R
    FASEB JOURNAL, 1997, 11 (03): : 3423 - 3423
  • [8] Whole-body pediatric phantoms for calibration of dual-energy X-ray absorptiometry.
    Shypailo, RJ
    Joo, JK
    Ellis, KJ
    FASEB JOURNAL, 1996, 10 (03): : 4496 - 4496
  • [9] WHOLE-BODY DUAL-ENERGY X-RAY ABSORPTIOMETRY IN CHILDREN, ADOLESCENTS AND YOUNG-ADULTS
    BRAILLON, PM
    COCHAT, P
    GIRAUD, S
    SALLE, BL
    DELMAS, PD
    MEUNIER, PJ
    JOURNAL OF BONE AND MINERAL RESEARCH, 1993, 8 : S291 - S291
  • [10] Effects of metal implants on whole-body dual-energy x-ray absorptiometry measurements of bone mineral content and body composition
    Giangregorio, LM
    Webber, CE
    CANADIAN ASSOCIATION OF RADIOLOGISTS JOURNAL-JOURNAL DE L ASSOCIATION CANADIENNE DES RADIOLOGISTES, 2003, 54 (05): : 305 - 309