SERVOCONTROLLED GENERATOR TO MEASURE RESPIRATORY IMPEDANCE FROM 0.25 TO 26 HZ IN VENTILATED PATIENTS AT DIFFERENT PEEP LEVELS

被引:30
作者
FARRE, R [1 ]
FERRER, M [1 ]
ROTGER, M [1 ]
NAVAJAS, D [1 ]
机构
[1] HOSP CLIN BARCELONA,SERV PNEUMOL,BARCELONA,SPAIN
关键词
FORCED OSCILLATION; IMPEDANCE MEASUREMENTS; MECHANICAL VENTILATION; MONITORING; POSITIVE END-EXPIRATORY PRESSURE;
D O I
10.1183/09031936.95.08071222
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Assessing respiratory impedance (Zrs) in ventilated patients over a wide frequency band, ranging from breathing rates to typical forced oscillation frequencies, during end-expiratory pauses at different positive end-expiratory pressures (PEEP) is of potential interest to assess a patient's respiratory mechanics, Zn measurements under these conditions are not possible with the present variants of the forced oscillation technique, The aim of this work was to design a forced oscillation generator operating from spontaneous breathing frequencies whilst withstanding PEEP. To this end, we constructed a generator based on a servocontrolled loudspeaker, This allowed the loudspeaker cone to remain at its resting position regardless of the external PEEP applied, The system was optimized by using a mechanical analogue, The clinical applicability of the servocontrolled generator was assessed by measuring Zrs in mechanically-ventilated chronic obstructive pulmonary disease (COPD) patients during end-expiratory pauses at different transrespiratory pressures. The forced oscillation generator designed may be easily applicable in practice since it is small and light. The system is able to withstand transrespiratory pressures of up to 17 hPa and allows the application of forced oscillation of sufficient amplitude (>2 hPa peak-to-peak, 0.25-26 Hz) to obtain reliable respiratory resistance and reactance data. The servocontrolled generator permits the assessment of respiratory mechanics over a wide frequency band ranging from breathing frequencies to the most typical forced oscillation frequencies during end-expiratory pauses at PEEPs within the conventional range.
引用
收藏
页码:1222 / 1227
页数:6
相关论文
共 12 条
  • [1] AN IMPROVED FORCED OSCILLATORY ESTIMATION OF RESPIRATORY IMPEDANCE
    DAROCZY, B
    HANTOS, Z
    [J]. INTERNATIONAL JOURNAL OF BIO-MEDICAL COMPUTING, 1982, 13 (03): : 221 - 235
  • [2] Farre R, 1991, EUR RESPIR REV, V1, P132
  • [3] MODELING OF LOW-FREQUENCY PULMONARY IMPEDANCE IN DOGS
    HANTOS, Z
    DAROCZY, B
    CSENDES, T
    SUKI, B
    NAGY, S
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1990, 68 (03) : 849 - 860
  • [4] A GENERAL MATHEMATICAL-MODEL FOR RESPIRATORY DYNAMICS RELEVANT TO THE CLINICAL SETTING
    MARINI, JJ
    CROOKE, PS
    [J]. AMERICAN REVIEW OF RESPIRATORY DISEASE, 1993, 147 (01): : 14 - 24
  • [5] A NEW ESTIMATOR TO MINIMIZE THE ERROR DUE TO BREATHING IN THE MEASUREMENT OF RESPIRATORY IMPEDANCE
    NAVAJAS, D
    FARRE, R
    ROTGER, M
    PESLIN, R
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (12) : 1001 - 1005
  • [6] RESPIRATORY INPUT IMPEDANCE IN ANESTHETIZED PARALYZED PATIENTS
    NAVAJAS, D
    FARRE, R
    CANET, J
    ROTGER, M
    SANCHIS, J
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1990, 69 (04) : 1372 - 1379
  • [7] NAVAJAS D, 1994, EUR RESPIR REV, V4, P216
  • [8] PESLIN R, 1993, EUR RESPIR J, V6, P772
  • [9] AN ACTIVE FEEDBACK-SYSTEM FOR ISOTONIC STUDIES OF SMOOTH-MUSCLE
    SABOURIN, PJ
    KINGMA, YJ
    BOWES, KL
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1991, 38 (06) : 614 - 616
  • [10] DESIGN AND CALIBRATION OF A HIGH-FREQUENCY OSCILLATORY VENTILATOR
    SIMON, BA
    MITZNER, W
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1991, 38 (02) : 214 - 218