INITIAL EXPERIENCE WITH A NEW SINGLE CHAMBER, DUAL SENSOR RATE-RESPONSIVE PACEMAKER

被引:21
|
作者
CONNELLY, DT
机构
[1] Cardiothoracic Centre, Liverpool
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1993年 / 16卷 / 09期
关键词
SENSOR BLENDING; SENSOR CROSSING; DUAL SENSOR RATE RESPONSIVE PACEMAKER;
D O I
10.1111/j.1540-8159.1993.tb01818.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
In August 1991, a new single chamber pacemaker became available that utilizes information from two sensors, activity and stimulus-to-T wave (QT) interval. We are reporting on the first 90 implants in 21 centers. T wave sensing was adequate at implantation in 88/90 patients, with a safety margin of > 100% in 86/90. Activity sensing was adequate in all patients. The contribution of each sensor (sensor blending) is programmable for each patient. Of 75 patients assessed at 1 month after implant, three have been programmed to ''Activity-Only'' mode, and 72 to dual sensor mode. Of these, 18 have been programmed to ''QT < Activity,'' 48 to ''QT = Activity,'' and 6 to ''QT > Activity.'' Forty-five patients underwent exercise testing in dual sensor mode and a subgroup of 15 also underwent exercise testing in Activity-Only mode. The dual sensor mode produced a more gradual increase in pacing rate. Sensor Cross Checking(TM) satisfactorily prevented a sustained high pacing rate in tests of false-positive activity sensing (tapping, vibrating pacemaker, or static pressure). The maximum pacing rate on walking downstairs (94.2 +/- 7.2 ppm) was similar to that produced by walking upstairs (91.6 +/- 5.9 ppm). We conclude that initial assessment of this dual sensor, single chamber, rate responsive pacemaker confirms that the algorithm for combining data from two sensors functions satisfactorily. Dual sensor rate responsive pacing may offer significant advantages over single sensor devices, and further studies of this novel device are indicated.
引用
收藏
页码:1833 / 1841
页数:9
相关论文
共 50 条
  • [21] UNDESIRABLE MODE SWITCHING WITH A DUAL CHAMBER RATE RESPONSIVE PACEMAKER
    PITNEY, MR
    MAY, CD
    DAVIS, MJ
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (04): : 729 - 737
  • [22] Electrocautery-induced tachycardia in a rate-responsive pacemaker
    Wong, DT
    Middleton, W
    ANESTHESIOLOGY, 2001, 94 (04) : 710 - 711
  • [23] INCREASE IN CARDIAC-OUTPUT WITH RATE-RESPONSIVE PACEMAKER
    BEYERSDORF, F
    KREUZER, J
    HAPP, J
    ZEGELMAN, M
    SATTER, P
    ANNALS OF THORACIC SURGERY, 1986, 42 (02): : 201 - 205
  • [24] INITIAL EXPERIENCE WITH MODE SWITCHING IN A DUAL SENSOR, DUAL-CHAMBER PACEMAKER IN PATIENTS WITH PAROXYSMAL ATRIAL TACHYARRHYTHMIAS
    DENDULK, K
    DIJKMAN, B
    PIETERSE, M
    WELLENS, H
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (11): : 1900 - 1907
  • [25] PACEMAKER MEDIATED TACHYCARDIAS IN SINGLE CHAMBER RATE RESPONSIVE PACING
    LAU, CP
    TAI, YT
    FONG, PC
    CHENG, CH
    CHUNG, FLW
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (12): : 1575 - 1579
  • [26] A RATE-RESPONSIVE PACEMAKER THAT PHYSIOLOGICALLY REDUCES PACING RATES AT REST
    MORRISTHURGOOD, J
    CHIANG, CM
    ROCHELLE, J
    STEINHAUS, B
    ILSLEY, C
    PAUL, V
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (11): : 1928 - 1932
  • [27] FUNCTION OF THE ACTIVITRAX RATE-RESPONSIVE PACEMAKER DURING TRAVEL BY AIR
    TOFF, WD
    LEEKS, C
    JOY, MD
    BENNETT, G
    CAMM, AJ
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1987, 10 (03): : 753 - 753
  • [28] PREGNANCY AND CESAREAN-SECTION IN A PATIENT WITH A RATE-RESPONSIVE PACEMAKER
    ANDERSEN, C
    OXHOJ, H
    ARNSBO, P
    LYBECKER, H
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (02): : 386 - 391
  • [29] TRANSILIAC VEIN APPROACH TO A RATE-RESPONSIVE PERMANENT PACEMAKER IMPLANTATION
    ANTONELLI, D
    FREEDBERG, NA
    ROSENFELD, T
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (08): : 1751 - 1752
  • [30] Monitor-induced tachycardia in a patient with a rate-responsive pacemaker
    Houtman, S
    Rinia, M
    Kalkman, C
    ANAESTHESIA, 2006, 61 (04) : 399 - 401