Artificial Urinary Sphincter: Long-Term Results and Patient Satisfaction

被引:50
作者
Montague, Drogo K. [1 ]
机构
[1] Cleveland Clin, Ctr Genitourinary Reconstruct, Glickman Urol & Kidney Inst, Dept Urol, 9500 Euclid Ave, Cleveland, OH 44195 USA
关键词
D O I
10.1155/2012/835290
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
The published evidence concerning the safety, efficacy, and patient satisfaction for implantation of the current model of the artificial urinary sphincter (AS 800) in men with post prostatectomy urinary incontinence was the objective of this review. A Pub Med English language literature search from 1995 to 2011 was performed. A majority of men who undergo AUS implantation for post prostatectomy urinary incontinence achieve satisfactory results (0 to 1 pad per day). Infection rates range from 0.46 to 7%, cuff erosion rates range from 3.8 to 10%, and urethral atrophy ranges from 9.6 to 11.4%. Kaplan-Meier 5 year projections for freedom from any reoperation were 50% for a small series and 79.4% for a larger series. Kaplan-Meier projections for freedom from mechanical failure were 79% at 5 years and 72% at 10 years. In another series 10 year projections for freedom from mechanical failure were 64%. Although the artificial urinary sphincter (AUS) is the gold standard for the treatment of this disorder, most men will continue to need at least one pad per day for protection, and they are subject to a significant chance of future AUS revision or replacement.
引用
收藏
页数:4
相关论文
共 24 条
[1]   Revision rate after artificial urinary sphincter implantation for incontinence after radical prostatectomy: Actuarial analysis [J].
Clemens, JQ ;
Schuster, TG ;
Konnak, JW ;
McGuire, EJ ;
Faerber, GJ .
JOURNAL OF UROLOGY, 2001, 166 (04) :1372-1375
[2]   Artificial urinary sphincter for post-prostatectomy incontinence in men who had prior radiotherapy: A risk and outcome analysis [J].
Gomha, MA ;
Boone, TB .
JOURNAL OF UROLOGY, 2002, 167 (02) :591-596
[3]   Artificial urinary sphincter for post-radical prostatectomy urinary incontinence: Long-term subjective results [J].
Gousse, AE ;
Madjar, S ;
Lambert, MM ;
Fishman, IJ .
JOURNAL OF UROLOGY, 2001, 166 (05) :1755-1758
[4]   Transcorporal artificial urinary sphincter cuff placement in cases requiring revision for erosion and urethral atrophy [J].
Guralnick, ML ;
Miller, E ;
Toh, KL ;
Webster, GD .
JOURNAL OF UROLOGY, 2002, 167 (05) :2075-2078
[5]   Quality of life and continence assessment of the artificial urinary sphincter in men with minimum 3.5 years of followup [J].
Haab, F ;
Trockman, BA ;
Zimmern, PE ;
Leach, GE .
JOURNAL OF UROLOGY, 1997, 158 (02) :435-439
[6]   Surgical Treatment of Stress Incontinence in Men [J].
Herschom, Sender ;
Bruschini, Homero ;
Comiter, Craig ;
Grise, Philippe ;
Hanus, Tomas ;
Kirschner-Hermanns, Ruth ;
Abrams, Paul .
NEUROUROLOGY AND URODYNAMICS, 2010, 29 (01) :179-190
[7]   Long-term durability and functional outcomes among patients with artificial urinary sphincters: A 10-year retrospective review from the University of Michigan [J].
Kim, Simon P. ;
Sarmast, Zubair ;
Daignault, Stephanie ;
Faerber, Gary J. ;
McGuire, Edward J. ;
Latini, Jerilyn M. .
JOURNAL OF UROLOGY, 2008, 179 (05) :1912-1916
[8]   Long-term experience with the double-cuff AMS 800 artificial urinary sphincter [J].
Kowalczyk, JJ ;
Spicer, DL ;
Mulcahy, JJ .
UROLOGY, 1996, 47 (06) :895-897
[9]   13 years of experience with artificial urinary sphincter implantation at Baylor College of Medicine [J].
Lai, H. Henry ;
Hsu, Elias I. ;
Teh, Bin S. ;
Butler, E. Brian ;
Boone, Timothy B. .
JOURNAL OF UROLOGY, 2007, 177 (03) :1021-1025
[10]   Post-prostatectomy incontinence and the artificial urinary sphincter: A long-term study of patient satisfaction and criteria for success [J].
Litwiller, SE ;
Kim, KB ;
Fone, PD ;
White, RWD ;
Stone, AR .
JOURNAL OF UROLOGY, 1996, 156 (06) :1975-1980