LINE-SHAPE DISTORTIONS IN MISALIGNED CUBE CORNER INTERFEROMETERS

被引:49
作者
KAUPPINEN, J
SAARINEN, P
机构
[1] Department of Applied Physics, University of Turku, Turku
来源
APPLIED OPTICS | 1992年 / 31卷 / 01期
关键词
FOURIER TRANSFORM INFRARED; LINE-SHAPE DISTORTIONS; CUBE CORNER; INTERFEROMETERS; MICHELSON INTERFEROMETER; MISALIGNMENT;
D O I
10.1364/AO.31.000069
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Line-shape distortions caused by the misalignment of the moving cube mirror in Fourier transform spectrometers have been described. A method of studying and correcting these distortions is presented. By using this method we can estimate the accuracy of the line position, which is especially important in high-resolution Fourier transform spectroscopy. The method is verified in simulations, and in practice it has been used to align the Oulu Fourier transform spectrometer.
引用
收藏
页码:69 / 74
页数:6
相关论文
共 6 条
[1]   CORRECTING ERRORS IN OPTICAL-PATH DIFFERENCE IN FOURIER SPECTROSCOPY - NEW ACCURATE METHOD [J].
KAUPPINEN, J ;
KARKKAINEN, T ;
KYRO, E .
APPLIED OPTICS, 1978, 17 (10) :1587-1594
[2]   DOUBLE-BEAM HIGH-RESOLUTION FOURIER SPECTROMETER FOR FAR INFRARED [J].
KAUPPINEN, J .
APPLIED OPTICS, 1975, 14 (08) :1987-1992
[3]   LARGE APERTURE CUBE CORNER INTERFEROMETER WITH A RESOLUTION OF 0.001 CM-1 [J].
KAUPPINEN, J ;
HORNEMAN, VM .
APPLIED OPTICS, 1991, 30 (18) :2575-2578
[4]   WORKING RESOLUTION OF 0.010 CM-1 BETWEEN 20 CM-1 AND 1200 CM-1 BY A FOURIER SPECTROMETER [J].
KAUPPINEN, J .
APPLIED OPTICS, 1979, 18 (11) :1788-1796
[5]  
KAUPPINEN J, 1985, 9TH C HIGH RES SPECT