NONIONIC CONTRAST-MEDIA - A COMPARISON OF IODINE DELIVERY RATES DURING MANUAL INJECTION ANGIOGRAPHY

被引:15
|
作者
HUGHES, PM
BISSET, R
机构
[1] N MANCHESTER GRP HOSP,DEPT RADIOL,MANCHESTER,ENGLAND
[2] HOPE HOSP,DEPT RADIOL,SALFORD M6 8HD,LANCS,ENGLAND
关键词
CONTRAST MEDIA; INJECTION RATE; ARTERIOGRAPHY;
D O I
10.1259/0007-1285-64-761-417
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Iodine delivery rates (IDR) of five commonly used non-ionic contrast media were determined at room temperature (24-degrees-C) and body temperature (37-degrees-C). Contrast media of strength 300 mgI/ml were also evaluated at 50% dilution (150 mgI/ml) with N-saline. Iodine delivery differed significantly (p < 0.005) between samples at room temperature: Omnipaque 350 (1163 mg/s) < Niopam 370 (1311 mg/s) < Omnipaque 300 (1422 mg/s) < Niopam 300 (1635 mg/s) and Ultravist 300 (1636 mg/s). Niopam 300 and Ultravist 300 delivered 41% more iodine per second than Omnipaque 350 at room temperature. Similar differences were identified at body temperature, while delivery of individual media was on average 23.5% greater than at room temperature. No significant differences between iodine delivery rates of diluted media at room temperature or body temperature was identified. The results demonstrate that iodine delivery and hence vascular opacification are better achieved during hand-injection arteriography by using relatively low viscosity media such as Niopam 300 or Ultravist 300. In digital subtraction arteriography all 300 strength contrast media diluted to 150 strength are equally effective.
引用
收藏
页码:417 / 419
页数:3
相关论文
共 50 条
  • [31] THROMBOGENICITY OF IONIC AND NONIONIC CONTRAST-MEDIA TESTED IN A LASER-INDUCED RAT THROMBOSIS MODEL
    AGUEJOUF, O
    DOUTREMEPUICH, F
    AZOUGAGH, OF
    DOUTREMEPUICH, C
    THROMBOSIS RESEARCH, 1995, 77 (03) : 259 - 269
  • [32] EFFECTS OF VASOCONSTRICTORS ON RABBIT CORONARY-ARTERIES EXPOSED TO ISOTONIC IONIC AND NONIONIC CONTRAST-MEDIA
    KARSTOFT, J
    BAATH, L
    JANSEN, I
    EDVINSSON, L
    INVESTIGATIVE RADIOLOGY, 1993, 28 (10) : 911 - 916
  • [33] NONIONIC CONTRAST-MEDIA IN PEDIATRIC CT - A COMPARATIVE-STUDY OF INTRAVENOUS USE OF IOPENTOL AND IOHEXOL
    WIKLUND, LM
    BROLIN, I
    SJOHOLM, K
    ACTA RADIOLOGICA, 1994, 35 (02) : 186 - 190
  • [34] CONTRAST-MEDIA LEAKAGE TO THE CEREBROSPINAL-FLUID AFTER INTRAVENOUS-INJECTION - COMPARISON OF STROKE PATIENTS AND CONTROLS
    LARSEN, JL
    SKOUEN, JS
    VIK, H
    ACTA RADIOLOGICA, 1995, 36 (04) : 440 - 447
  • [35] RENAL STREAK ARTIFACT DURING CONTRAST-ENHANCED CT - COMPARISON OF LOW VERSUS HIGH OSMOLALITY CONTRAST-MEDIA
    SUSSMAN, SK
    ILLESCAS, FF
    OPALACZ, JP
    YIRGA, P
    FOLEY, LC
    ABDOMINAL IMAGING, 1993, 18 (02): : 180 - 185
  • [36] VASOCONSTRICTION OF ISOLATED ARTERIES INDUCED BY ANGIOGRAPHIC CONTRAST-MEDIA - A COMPARISON OF IONIC AND NON-IONIC CONTRAST-MEDIA ISO-OSMOLAR WITH PLASMA
    KARSTOFT, J
    BAATH, L
    JANSEN, I
    EDVINSSON, L
    ACTA RADIOLOGICA, 1995, 36 (03) : 312 - 316
  • [37] Renal CT after intravenous injection of nonionic dimeric or nonionic monomeric contrast media in healthy volunteers
    Rasmussen, F
    Lindequist, S
    Nielsen, SM
    Bjartveit, K
    ACTA RADIOLOGICA, 1997, 38 (01) : 61 - 67
  • [38] SELECTIVE INJECTION OF CONTRAST-MEDIA - INFLAMMATORY EFFECTS ON RABBIT FALLOPIAN-TUBES
    THURMOND, AS
    HEDGPETH, PL
    SCANLAN, RM
    RADIOLOGY, 1991, 180 (01) : 97 - 99
  • [39] ACTIVATION OF PLATELETS BY LOW-OSMOLAR CONTRAST-MEDIA - DIFFERENTIAL-EFFECTS OF IONIC AND NONIONIC AGENTS
    HARDEMAN, MR
    KONIJNENBERG, A
    STURK, A
    REEKERS, JA
    RADIOLOGY, 1994, 192 (02) : 563 - 566
  • [40] Optimizing Contrast Media Injection Protocols in State-of-the Art Computed Tomographic Angiography
    Lell, Michael M.
    Jost, Gregor
    Korporaal, Johannes Georg
    Mahnken, Andreas Horst
    Flohr, Thomas G.
    Uder, Michael
    Pietsch, Hubertus
    INVESTIGATIVE RADIOLOGY, 2015, 50 (03) : 161 - 167