ONLINE FERMENTATION PROCESS MONITORING OF CARBOHYDRATES AND ETHANOL USING TANGENTIAL FLOW FILTRATION AND COLUMN LIQUID-CHROMATOGRAPHY

被引:54
作者
BUTTLER, TA
JOHANSSON, KAJ
GORTON, LGO
MARKOVARGA, GA
机构
[1] Department of Analytical Chemistry, University of Lund, S-221 00 Lund
关键词
D O I
10.1021/ac00067a014
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This paper describes a fully automated on-line system for the unattended monitoring of bioprocesses. Removal of solids and macromolecules from fermentation broth is achieved with a tangential flow filtration unit (Waters Filter/Acquisition Module). The obtained filtrate is transported in a flow system to a cleanup step, based on solid-phase extraction, followed by introduction into a column liquid chromatographic system. Carbohydrates are then detected by refractive index and ethanol content using a laboratory-made selective amperometric alcohol biosensor. Data acquisition and processing, as well as switching of the system valves, are performed by a computer. The on-line system was used for continuous monitoring of five carbohydrates and ethanol during 24-h ethanol fermentations. Waste water from the Swedish pulp industry was used as the technical substrate. Typical RSD values of the entire setup were on the order of 5-6% over a 24-h period of operation.
引用
收藏
页码:2628 / 2636
页数:9
相关论文
共 67 条
  • [11] A CONTINUOUS METHOD FOR MONITORING AND CONTROLLING FERMENTATIONS - USING AN AUTOMATED HPLC SYSTEM
    DINWOODIE, RC
    MEHNERT, DW
    [J]. BIOTECHNOLOGY AND BIOENGINEERING, 1985, 27 (07) : 1060 - 1062
  • [12] 2-STAGE STEAM PRETREATMENT OF WILLOW FOR INCREASED PENTOSE YIELD
    EKLUND, R
    GALBE, M
    ZACCHI, G
    [J]. JOURNAL OF WOOD CHEMISTRY AND TECHNOLOGY, 1988, 8 (03) : 379 - 392
  • [13] EKLUND R, 1988, 8 INT S ALC FUELS TO, P101
  • [14] ELMROTH I, 1992, BIOTECHNOL BIOENG, V35, P787
  • [15] ONLINE MONITORING AND CONTROL OF FERMENTATION PROCESSES BY FLOW-INJECTION ANALYSIS
    FORMAN, LW
    THOMAS, BD
    JACOBSON, FS
    [J]. ANALYTICA CHIMICA ACTA, 1991, 249 (01) : 101 - 111
  • [16] A FLOW-INJECTION ANALYSIS SYSTEM FOR FERMENTATION MONITORING AND CONTROL
    GARN, M
    GISIN, M
    THOMMEN, C
    CEVEY, P
    [J]. BIOTECHNOLOGY AND BIOENGINEERING, 1989, 34 (04) : 423 - 428
  • [17] STABILIZATION OF ANALYTICAL ENZYMES USING A NOVEL POLYMER CARBOHYDRATE SYSTEM AND THE PRODUCTION OF A STABILIZED, SINGLE REAGENT FOR ALCOHOL ANALYSIS
    GIBSON, TD
    HIGGINS, IJ
    WOODWARD, JR
    [J]. ANALYST, 1992, 117 (08) : 1293 - 1297
  • [18] AMPEROMETRIC BIOSENSORS BASED ON AN APPARENT DIRECT ELECTRON-TRANSFER BETWEEN ELECTRODES AND IMMOBILIZED PEROXIDASES - PLENARY LECTURE
    GORTON, L
    JONSSONPETTERSSON, G
    CSOREGI, E
    JOHANSSON, K
    DOMINGUEZ, E
    MARKOVARGA, G
    [J]. ANALYST, 1992, 117 (08) : 1235 - 1241
  • [19] SELECTIVE DETECTION IN FLOW-ANALYSIS BASED ON THE COMBINATION OF IMMOBILIZED ENZYMES AND CHEMICALLY MODIFIED ELECTRODES
    GORTON, L
    CSOREGI, E
    DOMINGUEZ, E
    EMNEUS, J
    JONSSONPETTERSSON, G
    MARKOVARGA, G
    PERSSON, B
    [J]. ANALYTICA CHIMICA ACTA, 1991, 250 (01) : 203 - 248
  • [20] GORTON L, 1993, BIOELECTROANALYSIS, V2, P33