ELECTRODE RADIUS PREDICTS LESION RADIUS DURING RADIOFREQUENCY ENERGY HEATING - VALIDATION OF A PROPOSED THERMODYNAMIC MODEL

被引:141
作者
HAINES, DE
WATSON, DD
VEROW, AF
机构
[1] UNIV VIRGINIA, MED CTR, SCH MED, DEPT RADIOL, CHARLOTTESVILLE, VA 22901 USA
[2] UNIV VIRGINIA, MED CTR, SCH MED, DEPT INTERNAL MED, DIV CARDIOL, CHARLOTTESVILLE, VA 22901 USA
关键词
arrhythmia; catheter ablation; radiofrequency energy; thermodynamic model;
D O I
10.1161/01.RES.67.1.124
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Myocardial heating by transcatheter delivery of radiofrequency (RF) energy has been proposed as an effective means of arrhythmia ablation. A thermodynamic model describing the radial temperature gradient at steady state during RF-induced heating is proposed. If one assumes that RF power output is adjusted to maintain a constant electrode-tissue interface temperature at all times, then this thermodynamic model predicts that the radius of the RF-induced lesion will be directly proportional to the electrode radius. A total of 76 RF-induced lesions were created in a model of isolated canine right ventricualr free wall perfused and superfused with oxygenated Krebs-Henseleit buffer. Electrode radius was varied between 0.75 and 2.25 mm. RF energy (500 kHz) was delivered for 90 seconds, and the power output was adjusted to maintain a constant electrode-tissue interface temperature of 60°C. A strong linear correlation was observed between electrode radius and lesion radius in two dimensions: transverse (p = 0.0001, r = 0.85) and transmural (p = 0.0001, r = 0.89). With these data, the temperature correlation with irreversible myocardial injury in this model was calculated at 46.6-48.8°C. Therefore, the proposed thermodynamic model closely predicts the observed relation between electrode radius and lesion size during RF myocardial heating.
引用
收藏
页码:124 / 129
页数:6
相关论文
共 26 条
[21]   CATHETER ABLATION OF THE ATRIOVENTRICULAR JUNCTION WITH RADIOFREQUENCY ENERGY [J].
LANGBERG, JJ ;
CHIN, MC ;
ROSENQVIST, M ;
COCKRELL, J ;
DULLET, N ;
VANHARE, G ;
GRIFFIN, JC ;
SCHEINMAN, MM .
CIRCULATION, 1989, 80 (06) :1527-1535
[22]  
MICHAELSON SM, 1987, BIOL EFFECTS HLTH IM, pCH7
[23]   CATHETER ABLATION FOR TREATMENT OF TACHYARRHYTHMIAS - PRESENT ROLE AND POTENTIAL PROMISE [J].
SCHEINMAN, MM ;
DAVIS, JC .
CIRCULATION, 1986, 73 (01) :10-13
[24]   CATHETER-INDUCED ABLATION OF THE ATRIOVENTRICULAR JUNCTION TO CONTROL REFRACTORY SUPRA-VENTRICULAR ARRHYTHMIAS [J].
SCHEINMAN, MM ;
MORADY, F ;
HESS, DS ;
GONZALEZ, R .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1982, 248 (07) :851-855
[25]  
SEARS FW, 1955, U PHYSICS, pCH17
[26]   TEMPERATURE DISTRIBUTIONS FROM INTERSTITIAL RF ELECTRODE HYPERTHERMIA SYSTEMS - THEORETICAL PREDICTIONS [J].
STROHBEHN, JW .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1983, 9 (11) :1655-1667