Evaluation of multiple factors affecting normal brain dose in single-isocenter multiple target radiosurgery

被引:1
|
作者
Yuan, Yu [1 ]
Thomas, Evan M. [1 ]
Clark, Grant A. [1 ]
Markert, James M. [2 ]
Fiveash, John B. [1 ]
Popple, Richard A. [1 ]
机构
[1] Univ Alabama Birmingham, Dept Radiat Oncol, 1720 2nd Ave South, Birmingham, AL 35294 USA
[2] Univ Alabama Birmingham, Dept Neurosurg, Birmingham, AL USA
来源
JOURNAL OF RADIOSURGERY AND SBRT | 2018年 / 5卷 / 02期
关键词
radiosurgery; VMAT; linac; SRS; brain metastases;
D O I
暂无
中图分类号
R61 [外科手术学];
学科分类号
摘要
We investigated the effects of multiple planning factors on normal brain dose for single-isocenter VMAT stereotactic radiosurgery (SRS). Ten patients were retrospectively planned using a standardized objective function and all 16 combinations of 2 versus 4 arcs, collimator angle 45 degrees versus selected per beam to minimize area of normal brain exposed in the beams-eye-view, fixed jaw versus following the trailing MLC leaf, and a 2 Gy mean dose objective for healthy brain versus no low dose objective. Limiting the normal brain mean dose in the optimization objective function significantly reduced the low dose spill into the normal brain without changing target coverage. Jaw tracking and appropriate selection of collimator also reduced the low dose volume, but to a lesser extent. To reduce low dose spill into normal brain for single isocenter VMAT radiosurgery of multiple targets, it is important to incorporate a limit on low dose spill into the objective function. This study has implications beyond single-isocenter VMAT radiosurgery. When comparing different inverse-planned treatment techniques, metrics that are important for evaluation of plan quality must be included the objective function.
引用
收藏
页码:131 / 144
页数:14
相关论文
共 50 条
  • [21] Physics considerations for single-isocenter, volumetric modulated arc radiosurgery for treatment of multiple intracranial targets
    Stanhope, Carl
    Chang, Zheng
    Wang, Zhiheng
    Yin, Fang-Fang
    Kim, Grace
    Salama, Joseph K.
    Kirkpatrick, John
    Adamson, Justus
    PRACTICAL RADIATION ONCOLOGY, 2016, 6 (03) : 207 - 213
  • [22] Local recurrence and radionecrosis after single-isocenter multiple targets stereotactic radiotherapy for brain metastases
    Crouzen, J. A.
    Zindler, J. D.
    Mast, M. E.
    Kleijnen, J. J. E.
    Versluis, M. C.
    Hashimzadah, M.
    Kiderlen, M.
    van Zyp, N. C. M. G. van der Voort
    Broekman, M. L. D.
    Petoukhova, A. L.
    SCIENTIFIC REPORTS, 2025, 15 (01):
  • [23] Single-Isocenter Linac-Based Radiosurgery for Brain Metastases with Coplanar Arcs: A Dosimetric and Clinical Analysis
    Faccenda, Valeria
    Panizza, Denis
    Pisoni, Valerio
    Trivellato, Sara
    Daniotti, Martina Camilla
    Bianchi, Sofia Paola
    De Ponti, Elena
    Arcangeli, Stefano
    CANCERS, 2023, 15 (18)
  • [24] Robustness of single-isocenter multiple-metastasis stereotactic radiosurgery end-to-end testing across institutions
    Saenz, Daniel
    Papanikolaou, Niko
    Zoros, Emmanouil
    Pappas, Evangelos
    Reiner, Michael
    Chew, Lip Teck
    Lim, Hooi Yin
    Hancock, Sam
    Nebelsky, Alex
    Njeh, Christopher
    Anagnostopoulos, Georgios
    JOURNAL OF RADIOSURGERY AND SBRT, 2020, 7 (03): : 223 - 232
  • [25] Commissioning cranial single-isocenter multi-target radiosurgery for the Versa HD
    Knill, Cory
    Sandhu, Raminder
    Halford, Robert
    Snyder, Michael
    Seymour, Zachary
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (04): : 108 - 114
  • [26] The effect of MLC leaf width in single-isocenter multi-target radiosurgery with volumetric modulated arc therapy
    Abisheva, Zhanerke
    Floyd, Scott R.
    Salama, Joseph K.
    Kirkpatrick, John
    Yin, Fang-Fang
    Moravan, Michael J.
    Giles, William
    Adamson, Justus
    JOURNAL OF RADIOSURGERY AND SBRT, 2019, 6 (02): : 131 - 138
  • [27] How Does the Number of Brain Metastases Correlate With Normal Brain Exposure in Single-Isocenter Multitarget Multifraction Stereotactic Radiosurgery
    Zheng, Dandan
    Yoon, Jihyung
    Jung, Hyunuk
    Lemus, Olga Maria Dona
    Gou, Lang
    Zhou, Yuwei
    Usuki, Kenneth Y.
    Hardy, Sara
    Milano, Michael T.
    ADVANCES IN RADIATION ONCOLOGY, 2024, 9 (06)
  • [28] Single-Isocenter Volumetric Modulated Arc Therapy vs. CyberKnife M6 for the Stereotactic Radiosurgery of Multiple Brain Metastases
    El Shafie, Rami A.
    Tonndorf-Martini, Eric
    Schmitt, Daniela
    Celik, Aylin
    Weber, Dorothea
    Lang, Kristin
    Koenig, Laila
    Hoene, Simon
    Forster, Tobias
    von Nettelbladt, Bastian
    Adeberg, Sebastian
    Debus, Jurgen
    Rieken, Stefan
    Bernhardt, Denise
    FRONTIERS IN ONCOLOGY, 2020, 10
  • [29] Radiobiological evaluation considering setup error on single-isocenter irradiation in stereotactic radiosurgery
    Nakano, Hisashi
    Tanabe, Satoshi
    Sasamoto, Ryuta
    Takizawa, Takeshi
    Utsunomiya, Satoru
    Sakai, Madoka
    Nakano, Toshimichi
    Ohta, Atsushi
    Kaidu, Motoki
    Ishikawa, Hiroyuki
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (07): : 266 - 275
  • [30] Single-Isocenter Volumetric-Modulated Arc Radiosurgery for Non-Small-Cell Lung Cancer Patients with Multiple Brain Metastases
    Uto, M.
    Katagiri, T.
    Takehana, K.
    Ogura, K.
    Mizowaki, T.
    JOURNAL OF THORACIC ONCOLOGY, 2018, 13 (10) : S702 - S703