The aim of this study is to derive the non-uniformity correction factor for the two therapy ionization chambers for the dose measurement near the brachytherapy source. The two ionization chambers of 0.6 cc and 0.1 cc volume were used. The measurement in air was performed for distances between 0.8 cm and 20 cm from the source in specially designed measurement jig. The non-uniformity correction factors were derived from the measured values. The experimentally derived factors were compared with the theoretically calculated non-uniformity correction factors and a close agreement was found between these two studies. The experimentally derived non-uniformity correction factor supports the anisotropic theory.
引用
收藏
页码:83 / 88
页数:6
相关论文
共 14 条
[1]  
[Anonymous], 1995, 51 AAPM
[2]   AN ANALYTIC THEORY OF THE POINT-SOURCE NONUNIFORMITY CORRECTION FACTOR FOR THICK-WALLED IONIZATION CHAMBERS IN PHOTON BEAMS [J].
BIELAJEW, AF .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (04) :517-538
[3]   CORRECTION FACTORS FOR THICK-WALLED IONIZATION CHAMBERS IN POINT-SOURCE PHOTON BEAMS [J].
BIELAJEW, AF .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (04) :501-516
[4]  
IAEA International Atomic Energy Agency. IAEA, 2002, CAL PHOT BET RAY SOU, V64, DOI [10.1016/s0167-8140(02)82497-5, DOI 10.1016/S0167-8140(02)82497-5]
[5]  
ICRU, 1976, 24 ICRU
[6]   EFFECT OF FINITE SIZE OF IONIZATION CHAMBERS ON MEASUREMENTS OF SMALL PHOTON SOURCES [J].
KONDO, S ;
RANDOLPH, ML .
RADIATION RESEARCH, 1960, 13 (01) :37-60
[7]  
MAYNEORD WV, 1937, BRIT J RADIOL, V10, P365
[8]   A COMPARISON OF RADIAL DOSE FUNCTIONS FOR PD-103, I-125, SM-145, AM-241, YB-169, IR-192, AND CS-137 BRACHYTHERAPY SOURCES [J].
MEIGOONI, AS ;
NATH, R .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1992, 22 (05) :1125-1130
[9]   Study of scattered radiation for in-air calibration by a multiple-distance method using ionization chambers and an HDR 192Ir brachytherapy source [J].
Patel, NP ;
Majumdar, B ;
Vijayan, V .
BRITISH JOURNAL OF RADIOLOGY, 2006, 79 (940) :347-352
[10]  
Patel NP, 2004, J MED PHYS, V29, P254