Keep an Eye on the Intracranial Pressure, Thanks to the Optic Nerve Sheath Diameter

被引:2
作者
Weinmann, Alexandre Vagh [1 ]
Beaucreux, Charlotte [1 ]
Kearns, Kevin [1 ]
Dubost, Clement [1 ]
机构
[1] Begin Mil Hosp, Intens Care Unit, St Mande, France
关键词
Brain hemodynamics; cerebral Doppler; cerebral monitoring; coma; optic nerve sheath diameter;
D O I
10.4103/ijccm.IJCCM_446_17
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Neurological examination on sedated patients is challenging and no gold standard monitoring is currently available. We report the case of a patient deeply sedated for the management of acute respiratory distress syndrome secondary to pneumonia. Delay in awakening led to cerebral exploration by transcranial Doppler (TCD) and measurement of optic nerve sheath diameter (ONSD). Abnormal values of ONSD prompted immediate brain computed tomography scan that confirmed high intracranial pressure (ICP) due to multiple deep parenchymal hematomas. Despite raised ICP, values obtained by TCD were normal, thanks to the persistence of cerebral autoregulation. This case highlights the interest of combining noninvasive techniques for neurological examination, especially for sedated patients.
引用
收藏
页码:460 / 462
页数:3
相关论文
共 8 条
[1]   Cerebral perfusion pressure in head-injured patients: a noninvasive assessment using transcranial Doppler ultrasonography [J].
Czosnyka, M ;
Matta, BF ;
Smielewski, P ;
Kirkpatrick, PJ ;
Pickard, JD .
JOURNAL OF NEUROSURGERY, 1998, 88 (05) :802-808
[2]  
Dubost C, 2012, ANN FR ANESTH, V31, P125
[3]   Ultrasonography of optic nerve sheath diameter for detection of raised intracranial pressure: a systematic review and meta-analysis [J].
Dubourg, Julie ;
Javouhey, Etienne ;
Geeraerts, Thomas ;
Messerer, Mahmoud ;
Kassai, Behrouz .
INTENSIVE CARE MEDICINE, 2011, 37 (07) :1059-1068
[4]   Non-invasive assessment of intracranial pressure using ocular sonography in neurocritical care patients [J].
Geeraerts, Thomas ;
Merceron, Sybille ;
Benhamou, Dan ;
Vigue, Bernard ;
Duranteau, Jacques .
INTENSIVE CARE MEDICINE, 2008, 34 (11) :2062-2067
[5]   Daily interruption of sedative infusions in critically ill patients undergoing mechanical ventilation [J].
Kress, JP ;
Pohlman, AS ;
O'Connor, MF ;
Hall, JB .
NEW ENGLAND JOURNAL OF MEDICINE, 2000, 342 (20) :1471-1477
[6]   Differentiating midazolam over-sedation from neurological damage in the intensive care unit [J].
McKenzie, CA ;
McKinnon, W ;
Naughton, DP ;
Treacher, D ;
Davies, G ;
Phillips, GJ ;
Hilton, PJ .
CRITICAL CARE, 2005, 9 (01) :R32-R36
[7]   The CAMINO intracranial pressure device in clinical practice:: Reliability, handling characteristics and complications [J].
Münch, E ;
Weigel, R ;
Schmiedek, P ;
Schürer, L .
ACTA NEUROCHIRURGICA, 1998, 140 (11) :1113-+
[8]   How reliable is the Bispectral Index in critically ill patients? A prospective, comparative, single-blinded observer study [J].
Nasraway, SA ;
Wu, EC ;
Kelleher, RM ;
Yasuda, CM ;
Donnelly, AM .
CRITICAL CARE MEDICINE, 2002, 30 (07) :1483-1487