Measurement of back-scattered radiation from micro multileaf collimator into the beam monitor chamber from a dual energy linear accelerator

被引:1
作者
Muralidhar, K. R. [2 ]
Murthy, P. Narayana [1 ]
Sresty, N. V. N. M. [2 ]
Dixit, Pramod Kumar [2 ]
Kumar, Rajneesh [2 ]
Raju, A. K. [2 ]
机构
[1] Nagarjuna Univ, Dept Phys, Guntur, India
[2] Indo Amer Canc Inst & Res Ctr, Rd 14,Banjara Hills, Hyderabad 500034, Andhra Pradesh, India
关键词
Backscattered radiation; beam monitor chamber; micro multileaf collimator;
D O I
10.4103/0971-6203.33243
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Measurements designed to find the collimator backscatter into the beam monitor chamber from Micro Multileaf collimator of 6 MV photon beams of the Siemens Primus linear accelerator were made with the help of dose rate feedback control. The photons and electrons backscattered from the upper and lower secondary collimator jaws give rise to a significant increase in the ion charge measured by monitor chamber. This increase varies between the different accelerators. The output measurements were carried out in air at the isocenter. The effect of collimator backscatter was investigated by measuring the pulse width, number of beam pulses per monitor unit, monitor unit rate and dose for different mMLC openings. These measurements were made with and without dose rate feedback control, i.e., with constant electron beam current in the accelerator. Monitor unit rate (MU/min) was almost constant for all field sizes. The maximum variation between the open and the closed feedback control circuits was 2.5%. There was no difference in pulse width and negligible difference in pulse frequency. Maximum value of backscattered radiation from the micro Multileaf collimator into the beam monitor chamber was found to be 0.5%.
引用
收藏
页码:65 / 67
页数:3
相关论文
共 13 条
[1]   BACKSCATTER INTO THE BEAM MONITOR CHAMBER - IMPLICATIONS FOR DOSIMETRY OF ASYMMETRIC COLLIMATORS [J].
DUZENLI, C ;
MCCLEAN, B ;
FIELD, C .
MEDICAL PHYSICS, 1993, 20 (02) :363-367
[2]   THE EFFECT OF COLLIMATOR BACKSCATTER RADIATION ON PHOTON OUTPUT OF LINEAR ACCELERATORS [J].
HUANG, PH ;
CHU, J ;
BJARNGARD, BE .
MEDICAL PHYSICS, 1987, 14 (02) :268-269
[3]   HEAD SCATTER DATA FOR SEVERAL LINEAR ACCELERATORS (4-18MV) [J].
KASE, KR ;
SVENSSON, GK .
MEDICAL PHYSICS, 1986, 13 (04) :530-532
[4]  
Khan F.M., 2003, PHYS RAD THERAPY, V3rd ed
[6]   Measurement of backscatter to the monitor chamber of medical accelerators using target charge [J].
Lam, KL ;
Muthuswamy, MS ;
Ten Haken, RK .
MEDICAL PHYSICS, 1998, 25 (03) :334-338
[7]   OUTPUT FACTOR CONSTITUENTS OF A HIGH-ENERGY PHOTON-BEAM [J].
LUXTON, G ;
ASTRAHAN, MA .
MEDICAL PHYSICS, 1988, 15 (01) :88-91
[8]   CHARACTERISTICS OF AN 18 MV PHOTON-BEAM FROM A THERAC 20 MEDICAL LINEAR-ACCELERATOR [J].
PATTERSON, MS ;
SHRAGGE, PC .
MEDICAL PHYSICS, 1981, 8 (03) :312-318
[9]   Measurement of backscattered radiation from secondary collimator jaws into the beam monitor chamber from a dual energy linear accelerator [J].
Ravikumar, M ;
Ravichandran, R .
STRAHLENTHERAPIE UND ONKOLOGIE, 2001, 177 (12) :670-+
[10]   BACKSCATTER FROM HIGH ATOMIC NUMBER MATERIALS IN HIGH-ENERGY PHOTON BEAMS [J].
SCRIMGER, JW .
RADIOLOGY, 1977, 124 (03) :815-817