COMPLICATIONS OF CEREBROSPINAL-FLUID SHUNT ANTISIPHON DEVICES

被引:0
|
作者
DASILVA, MC [1 ]
DRAKE, JM [1 ]
机构
[1] UNIV TORONTO,HOSP SICK CHILDREN,DEPT NEUROSURG,DIV NEUROSURG,555 UNIV AVE,TORONTO M5G 1X8,ONTARIO,CANADA
关键词
HYDROCEPHALUS; CEREBROSPINAL FLUID SHUNTS; ANTISIPHON DEVICE; CHILDREN;
D O I
暂无
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The case histories of 38 children with implanted antisiphon devices (ASD) were retrospectively reviewed. Sixty-three percent suffered a shunt malfunction within 5 years of ASD implantation. The probability of shunt malfunction within 1 year was of 40%. Patients with myelomeningocele experienced the highest rate of malfunction. Age, order of ASD implantation (whether implanted primarily or secondarily), shunt pressure, and reason for ASD insertion did not significantly alter the incidence of malfunction. Fifty percent of the shunt complications were specifically related to ASD malfunction. Functional obstruction of the shunt system as determined by evidence of clinical shunt obstruction with documented patency on shunt flow testing occurred in 10 of 38 patients. No variable studied could be significantly related to the probability of developing an ASD-related complication. Functional obstruction of the ASD is a common and unpredictable complication of these devices.
引用
收藏
页码:304 / 309
页数:6
相关论文
共 50 条
  • [21] Micro devices for a cerebrospinal fluid (CSF) shunt system
    Yoon, HJ
    Jung, JM
    Jeong, JS
    Yang, SS
    SENSORS AND ACTUATORS A-PHYSICAL, 2004, 110 (1-3) : 68 - 76
  • [22] Cerebrospinal fluid shunt complications - A 10-year study of etiologies and cerebrospinal fluid characteristics
    Rad, MF
    Vahedi, P
    Shoeibi, A
    NEUROSURGERY QUARTERLY, 2005, 15 (01) : 1 - 4
  • [23] Cerebrospinal fluid shunt malfunctions: A reflective review
    Low, Sharon Y. Y.
    Kestle, John R. W.
    Walker, Marion L.
    Seow, Wan Tew
    CHILDS NERVOUS SYSTEM, 2023, 39 (10) : 2719 - 2728
  • [24] CEREBROSPINAL-FLUID SHUNT DYNAMICS IN PATIENTS WITH IDIOPATHIC ADULT HYDROCEPHALUS SYNDROME
    MALM, J
    KRISTENSEN, B
    FAGERLUND, M
    KOSKINEN, LO
    EKSTEDT, J
    JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1995, 58 (06) : 715 - 723
  • [25] INDICATORS OF HYPOXIA IN CEREBROSPINAL-FLUID OF HYDROCEPHALIC CHILDREN WITH SUSPECTED SHUNT MALFUNCTION
    CASTROGAGO, M
    RODRIGUEZSEGADE, S
    CAMINA, F
    BOLLAR, A
    RODRIGUEZNUNEZ, A
    CHILDS NERVOUS SYSTEM, 1993, 9 (05) : 275 - 277
  • [26] PERITONEAL CEREBROSPINAL-FLUID SHUNT INSERTION - A TECHNIQUE FOR PROTECTION OF THE ABDOMINAL CATHETER
    PIATT, JH
    JOURNAL OF NEUROSURGERY, 1995, 82 (02) : 305 - 306
  • [27] RISK OF INFECTION AFTER CEREBROSPINAL-FLUID SHUNT - AN ANALYSIS OF 884 FIRST-TIME SHUNTS
    BORGBJERG, BM
    GJERRIS, E
    ALBECK, MJ
    ACTA NEUROCHIRURGICA, 1995, 136 (1-2) : 1 - 7
  • [28] The cerebrospinal fluid shunt
    Puget, S
    ARCHIVES DE PEDIATRIE, 2005, 12 (02): : 224 - 227
  • [29] DETERMINATION OF CEREBROSPINAL-FLUID SHUNT OBSTRUCTION WITH MAGNETIC-RESONANCE PHASE IMAGING
    DRAKE, JM
    MARTIN, AJ
    HENKLEMAN, RM
    JOURNAL OF NEUROSURGERY, 1991, 75 (04) : 535 - 540
  • [30] CEREBROSPINAL-FLUID FLOW VOID IN CHILDREN
    HAYAKAWA, K
    KONISHI, Y
    KURIYAMA, M
    KONISHI, K
    MATSUDA, T
    NEURORADIOLOGY, 1993, 35 (06) : 443 - 446