AN EVALUATION OF 2 ACTIVATED CLOTTING TIME MONITORS DURING CARDIAC-SURGERY

被引:10
|
作者
REICH, DL
ZAHL, K
PERUCHO, MH
THYS, DM
机构
[1] From the Department of Anesthesiology, Mount Sinai Medical Center, New York, 10029-6574, NY, One Gustave L. Levy Place
来源
JOURNAL OF CLINICAL MONITORING | 1992年 / 8卷 / 01期
关键词
BLOOD; COAGULATION; HEPARIN;
D O I
10.1007/BF01618085
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The activated clotting time (ACT) is a commonly used method for assessing the degree of anticoagulation during cardiac surgery. Two automated ACT monitors were evaluated in 29 adult patients undergoing cardiac surgery. The HemoTec ACT monitor (HTC; HemoTec, Inc., Englewood, CO) was evaluated using 0.8 ml of whole blood in dual-chamber, high range kaolin cartridges. The Hemochron ACT monitor (HCH; International Technidyne, Inc, Edison, NJ) was evaluated using 2.0 ml of whole blood in glass tubes with diatomaceous earth activator. Following sternotomy, a coagulation profile consisting of HTC, HCH, and a partial thromboplastin time (PTT) was obtained (TO). Beef lung heparin was administered in 3 consecutive doses: 40 units/kg (T1, 80 units/kg (T2), and 180 units/kg (T3). Coagulation profiles were drawn 5 minutes after each dose. Following cardiopulmonary bypass (CPB), coagulation profiles were drawn 15 minutes (T4), 2 hours (T5), and 24 hours (T6) after the protamine dose. HTC and HCH ACT values differed significantly at T1, T2, and T4 (p < 0.001). In the pre-CPB period, the HCH ACT correlation with PTT divided by its control value (PTT/CTL) (r = 0.73) was significantly better (p = 0.02) than the correlation of HTC ACT with PTT/CTL (r = 0.41). In the post-CPB period, both HCH ACT (r = 0.45) and HTC ACT (r = 0.30) correlated weakly with PTT/CTL. In the bias analysis, the limits of agreement (of all HCH and HTC ACT values) showed that HTC ACT is between 61% and 133% of the HCH ACT value in 95% of determinations. These data contradict the results of a previous comparison of the two monitors, a study that suggested that HTC ACT overestimated anticoagulation at low heparin doses. HCH ACT correlated better than HTC ACT with the PTT values only in the pre-CPB period. The bias analysis indicates that even though the mean ACT values are usually close, there is considerable spread (-39 to +33%) when the 95% confidence intervals are calculated.
引用
收藏
页码:33 / 36
页数:4
相关论文
共 50 条
  • [1] Con: Activated Clotting Time Should Not Be Monitored During Heparinization for Vascular Surgery
    Wolo, Elizabeth
    Herman, Corey
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2018, 32 (03) : 1497 - 1499
  • [2] The activated clotting time in cardiac surgery: should Celite or kaolin be used?
    De Vries, Adrianus J.
    Lansink-Hartgring, Annemieke Oude
    Fernhout, Freek-Jan
    Huet, Rolf C. G.
    van den Heuvel, Edwin R.
    INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2017, 24 (04) : 549 - 554
  • [3] Anticoagulation monitoring during vascular surgery:: accuracy of the Hemochron® low range activated clotting time (ACT-LR)
    Tremey, B.
    Szekely, B.
    Schlumberger, S.
    Francois, D.
    Liu, N.
    Sievert, K.
    Fischler, M.
    BRITISH JOURNAL OF ANAESTHESIA, 2006, 97 (04) : 453 - 459
  • [4] The relationship between heparin level and activated clotting time in the adult cardiac surgery population
    FitzGerald, D. J.
    Patel, A.
    Body, S. C.
    Garvin, S.
    PERFUSION-UK, 2009, 24 (02): : 93 - 96
  • [5] Post-Neutralisation Activated Clotting Time and Postoperative Transfusions in Cardiac Surgery Outcome
    Zahid, Muhammad Arslan
    Ahmed, Syed Shabbir
    Yousuf, Muhammad Saad
    Wadho, Sadique Ali
    Akhtar, Muhammad Irfan
    Hamid, Mohammad
    JCPSP-JOURNAL OF THE COLLEGE OF PHYSICIANS AND SURGEONS PAKISTAN, 2024, 34 (11): : 1275 - 1280
  • [6] Pro: Activated Clotting Time Should Be Monitored During Heparinization For Vascular Surgery
    Goldhammer, Jordan E.
    Zimmerman, Darin
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2018, 32 (03) : 1494 - 1496
  • [7] A comparative analysis of four activated clotting time measurement devices in cardiac surgery with cardiopulmonary bypass
    Li, Han
    Serrick, Cyril
    Rao, Vivek
    Yip, Paul M.
    PERFUSION-UK, 2021, 36 (06): : 610 - 619
  • [8] CHARACTERIZATION OF PROTHROMBIN ACTIVATION DURING CARDIAC-SURGERY BY HEMOSTATIC MOLECULAR MARKERS
    SLAUGHTER, TF
    LEBLEU, TH
    DOUGLAS, JM
    LESLIE, JB
    PARKER, JK
    GREENBERG, CS
    ANESTHESIOLOGY, 1994, 80 (03) : 520 - 526
  • [9] Activated clotting time-guided heparinization during open AAA surgery: a pilot study
    Roosendaal, Liliane C.
    Hoebink, Max
    Wiersema, Arno M.
    Blankensteijn, Jan D.
    Jongkind, Vincent
    PILOT AND FEASIBILITY STUDIES, 2024, 10 (01)
  • [10] PREOPERATIVE PLATELETPHERESIS DOES NOT REDUCE BLOOD-LOSS DURING CARDIAC-SURGERY
    BOEY, SK
    ONG, BC
    DHARA, SS
    CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1993, 40 (09): : 844 - 850