THE EFFECTS OF NORMOTHERMIC AND HYPOTHERMIC CARDIOPULMONARY BYPASS ON DEFIBRILLATION ENERGY-REQUIREMENTS AND TRANSMYOCARDIAL IMPEDANCE - IMPLICATIONS FOR IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR IMPLANTATION

被引:1
作者
MARTIN, D
GARCIA, J
VALERI, CR
KHURI, SF
LIM, C
DOURSOUNIAN, M
机构
[1] VET ADM MED CTR W ROXBURY,DEPT CARDIOL,BOSTON,MA
[2] VET ADM MED CTR W ROXBURY,SURG SERV,BOSTON,MA
[3] BRIGHAM & WOMENS HOSP,DEPT MED,BOSTON,MA
[4] BRIGHAM & WOMENS HOSP,DEPT SURG,BOSTON,MA
[5] HARVARD UNIV,SCH MED,CAMBRIDGE,MA
[6] BOSTON UNIV,SCH MED,NAVAL BLOOD RES LAB,BOSTON,MA 02118
关键词
D O I
10.1016/S0022-5223(95)70324-1
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The influence of normothermic and hypothermic cardiopulmonary bypass on defibrillation energy requirements and transcardiac impedance is not well characterized. However, this relationship is of clinical importance during automatic defibrillator implantation done with concomitant cardiac surgery, and there is anecdotal information that criteria for successful implantation are harder to achieve after such operations. We studied the effect of controlled hypothermia on defibrillation energy requirements and transcardiac impedance in a canine model of cardiopulmonary bypass in which 26 animals underwent right atrial and femoral arterial cannulation, as well as continuous hemodynamic and intramyocardial temperature monitoring. The defibrillation energy requirements were evaluated at 60-minute intervals with an epicardial patch system, and transcardiac impedance was measured before and after the multiple inductions and terminations of ventricular fibrillation. In group 1 (n = 10) defibrillation energy requirements were evaluated immediately after initiation of cardiopulmonary bypass at 37 degrees C (T-0), after gradual cooling to 28 degrees C (T-1), and after rewarming to 37 degrees C (T-2), Group 2 (n = 16) comprised time controls that were identically instrumented and studied, but maintained at 37 degrees C throughout. Percent successful defibrillation was plotted against delivered energy, and the raw data fit by logistic regression. The energy at which 50% of shocks were successful (E(50)) was 3.23 +/- 0.89 joules at T-0, 5.12 +/- 1.85 joules at T-1, and 4.42 +/- 1.22 joules at T-2 in group 1; this was not significantly different from the corresponding group 2 E(50) values, which were 3.11 +/- 1.39 joules, 4.95 +/- 2.47 joules, and 5.59 +/- 3.18 joules, respectively. Both groups demonstrated a significant increase in E(50), during the first hour of cardiopulmonary bypass (mean increase from T-0 to T-1 was 1.89 joules in group 1 and 1.84 joules in group 2, p < 0.05). Transmyocardial impedance fell progressively during the group 2 experiments from 73.6 +/- 12.9 Omega at the beginning of the T-0 shock series to 61.4 +/- 8.9 Omega at the end of the T-2 shock series. A similar reduction in transmyocardial impedance was observed during the course of all the group 1 experiments; however, at the beginning of the T-1 shock series impedance was significantly elevated to 77.4 +/- 12.3 Omega (p < 0.05 compared with group 2 and with end T-1 in group 1). There was no relationship between defibrillation energy requirements and transcardiac impedance; there was also no correlation between either of these parameters and intramyocardial extracellular pH or left ventricular end diastolic pressure. We conclude that defibrillation energy requirements significantly and persistently increase during the first hour of cardiopulmonary bypass and the effect is temperature independent. Systemic hypothermia significantly and reversibly elevates transmyocardial impedance but multiple shocks: reduce this effect at 28 degrees C as they do at 37 degrees C.
引用
收藏
页码:981 / 988
页数:8
相关论文
共 32 条
  • [1] TEMPORAL STABILITY AND PRECISION OF VENTRICULAR DEFIBRILLATION THRESHOLD DATA
    BABBS, CF
    WHISTLER, SJ
    YIM, GKW
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY, 1978, 235 (05): : H553 - H558
  • [2] Dahl C F, 1976, Med Instrum, V10, P151
  • [3] DALZELL GWN, 1991, BRIT HEART J, V65, P311
  • [4] THE RELATIONSHIP BETWEEN SUCCESSFUL DEFIBRILLATION AND DELIVERED ENERGY IN OPEN-CHEST DOGS - REAPPRAISAL OF THE DEFIBRILLATION THRESHOLD CONCEPT
    DAVY, JM
    FAIN, ES
    DORIAN, P
    WINKLE, RA
    [J]. AMERICAN HEART JOURNAL, 1987, 113 (01) : 77 - 84
  • [5] PH-DEPENDENT EFFECTS OF LIDOCAINE ON DEFIBRILLATION ENERGY-REQUIREMENTS IN DOGS
    ECHT, DS
    CATO, EL
    COXE, DR
    [J]. CIRCULATION, 1989, 80 (04) : 1003 - 1009
  • [6] EVALUATION OF ANTIARRHYTHMIC DRUGS ON DEFIBRILLATION ENERGY-REQUIREMENTS IN DOGS - SODIUM-CHANNEL BLOCK AND ACTION-POTENTIAL PROLONGATION
    ECHT, DS
    BLACK, JN
    BARBEY, JT
    COXE, DR
    CATO, E
    [J]. CIRCULATION, 1989, 79 (05) : 1106 - 1117
  • [7] BIOELECTRICAL TISSUE RESISTANCE DURING VARIOUS METHODS OF MYOCARDIAL PRESERVATION
    GARRIDO, H
    SUEIRO, J
    RIVAS, J
    VILCHES, J
    ROMERO, JM
    GARRIDO, F
    [J]. ANNALS OF THORACIC SURGERY, 1983, 36 (02) : 143 - 151
  • [8] Geddes L A, 1975, Med Instrum, V9, P179
  • [9] THE DEFIBRILLATION SUCCESS RATE VERSUS ENERGY RELATIONSHIP .1. CURVE FITTING AND THE MOST EFFICIENT DEFIBRILLATION ENERGY
    GLINER, BE
    MURAKAWA, Y
    THAKOR, NV
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (03): : 326 - 338
  • [10] STRETCH-INDUCED ARRHYTHMIAS IN THE ISOLATED CANINE VENTRICLE - EVIDENCE FOR THE IMPORTANCE OF MECHANOELECTRICAL FEEDBACK
    HANSEN, DE
    CRAIG, CS
    HONDEGHEM, LM
    [J]. CIRCULATION, 1990, 81 (03) : 1094 - 1105