Interference screw versus Endoscrew fixation for anterior cruciate ligament reconstruction: A biomechanical comparative study in sawbones and porcine knees

被引:15
|
作者
Hung, Chu-Chih [1 ]
Chen, Wen-Chuan [1 ]
Yang, Chan-Tsung [1 ]
Cheng, Cheng-Kung [1 ,2 ]
Chen, Chih-Hwa [3 ]
Lai, Yu-Shu [1 ]
机构
[1] Natl Yang Ming Univ, Orthopaed Device Res Ctr, 155,Sect 2,Linong St,Shih Pai, Taipei 11221, Taiwan
[2] Natl Yang Ming Univ, Inst Biomed Engn, Taipei, Taiwan
[3] Taipei Med Univ Hosp, Taipei Med Univ, Coll Med, Sch Med,Dept Orthopaed & Traumatol, Taipei 11221, Taiwan
关键词
Anterior cruciate ligament reconstruction; Biomechanics; Knee; Orthopaedic fixation devices; Porcine model;
D O I
10.1016/j.jot.2014.02.001
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Interference screw fixation is one of the most common methods for ligament reconstruction. Although the advantages and clinical outcomes of this procedure have been widely reported, post-surgical complications often arise. The purpose of this study was to evaluate a new femoral fixation device, the Endoscrew, for anterior cruciate ligament (ACL) reconstruction. We performed a mechanical test in accordance with American Society for Testing and Materials (ASTM) standards and an in vitro biomechanical study. An axial pullout test was conducted to evaluate the mechanical properties of the new device and the interference screw when implanted in solid rigid polyurethane foam test blocks. The biomechanical test used porcine femora to evaluate the initial fixation strength between these two implants. The maximum pullout force of the interference screw group [722.05 +/- 130.49 N (N)] was significantly greater (p < 0.01) than the Endoscrew group (440.79 +/- 26.54 N) when implanted in polyurethane foam 320 kg/m(3) density. With polyurethane foam 160 kg/m(3) density, the maximum pullout forces were (242.61 +/- 37.36 N) (p < 0.001) and (99.33 +/- 30.01 N) for the interference screw group and Endoscrew group, respectively. In the in vitro mechanical study, the Endoscrew (646.39 +/- 72.38 N) required a significantly greater ultimate load prior to failure (p < 0.05) when compared with the interference screw (489.72 +/- 138.64 N). With regard to pullout stiffness, there was no statistically significant difference (p < 0.13) between the Endoscrew group (99.15 +/- 12.16 N/mm) and the interference screw group (87.96 +/- 11.12 N/mm). The cyclic stiffness was also not significantly different (p < 0.44) between the Endoscrew group (93.09 +/- 16.07 N/mm) and the interference screw group (85.78 +/- 14.76 N/mm). The axial pullout test showed that the strength of the Endoscrew was close to the fixation strength required for daily activities, but it is inappropriate for use in osteoporotic patients. The in vitro study showed that the Endoscrew has a superior initial mechanical fixation over the biodegradable interference screw. Copyright (C) 2014, The Authors. Published by Elsevier (Singapore) Pte Ltd.
引用
收藏
页码:82 / 90
页数:9
相关论文
共 50 条
  • [21] Comparison of femoral fixation methods for anterior cruciate ligament reconstruction with patellar tendon graft: a mechanical analysis in porcine knees
    Giuseppe Milano
    Pier Damiano Mulas
    Fabio Ziranu
    Laura Deriu
    Carlo Fabbriciani
    Knee Surgery, Sports Traumatology, Arthroscopy, 2007, 15 : 733 - 738
  • [22] Biomechanical study of a computer simulated reconstruction of the anterior cruciate ligament (ACL)
    Veselko, M
    Godler, I
    COMPUTERS IN BIOLOGY AND MEDICINE, 2000, 30 (06) : 299 - 309
  • [23] In Vitro Testing of 2 Adjustable-Loop Cortical Suspensory Fixation Systems Versus Interference Screw for Anterior Cruciate Ligament Reconstruction
    Garces, Gerardo L.
    Martel, Oscar
    Yanez, Alejandro
    Manchado-Herrera, Ignacio
    Motta, Luci M.
    ORTHOPAEDIC JOURNAL OF SPORTS MEDICINE, 2021, 9 (09)
  • [24] Biomechanical comparisons of current suspensory fixation devices for anterior cruciate ligament reconstruction
    Jin Cheng
    Siddhartha Venkata Paluvadi
    SungJae Lee
    SeungJin Yoo
    Eun-Kyoo Song
    Jong-Keun Seon
    International Orthopaedics, 2018, 42 : 1291 - 1296
  • [25] Initial fixation strength of interference nail fixation for anterior cruciate ligament reconstruction with patellar tendon graft (experimental study)
    Aydin, H
    Kerimoglu, S
    Çitlak, A
    Turhan, AU
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2004, 12 (02) : 94 - 97
  • [26] Double-bundle anterior cruciate ligament reconstruction using hamstring autografts and bioabsorbable interference screw fixation
    Jaervelae, Timo
    Moisala, Anna-Stina
    Sihvonen, Raine
    Jaervelae, Sally
    Kannus, Pekka
    Jaervinen, Markku
    AMERICAN JOURNAL OF SPORTS MEDICINE, 2008, 36 (02) : 290 - 297
  • [27] Biomechanical comparisons of current suspensory fixation devices for anterior cruciate ligament reconstruction
    Cheng, Jin
    Paluvadi, Siddhartha Venkata
    Lee, SungJae
    Yoo, SeungJin
    Song, Eun-Kyoo
    Seon, Jong-Keun
    INTERNATIONAL ORTHOPAEDICS, 2018, 42 (06) : 1291 - 1296
  • [28] Novel press-fit technique of patellar bone plug in anterior cruciate ligament reconstruction is comparable to interference screw fixation
    Haberli, Janosch
    Heilgemeir, Maximilian
    Valet, Sebastian
    Aiyangar, Ameet
    Overes, Tom
    Henle, Philipp
    Eggli, Stefan
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2022, 142 (08) : 1963 - 1970
  • [29] Initial fixation strength of interference nail fixation for anterior cruciate ligament reconstruction with patellar tendon graft (experimental study)
    H. Aydın
    S. Kerimoğlu
    A. Çıtlak
    A. U. Turhan
    Knee Surgery, Sports Traumatology, Arthroscopy, 2004, 12 : 94 - 97
  • [30] Tibial Fixation in Anterior Cruciate Ligament Reconstruction A Prospective Randomized Study Comparing Metal Interference Screw and Staples With a Centrally Placed Polyethylene Screw and Sheath
    De Wall, Mathew
    Scholes, Corey J.
    Patel, Sandeep
    Coolican, Myles R. J.
    Parker, David A.
    AMERICAN JOURNAL OF SPORTS MEDICINE, 2011, 39 (09) : 1858 - 1864