PATIENT DOSE VALUES IN INTERVENTIONAL RADIOLOGY

被引:139
作者
VANO, E [1 ]
GONZALEZ, L [1 ]
FERNANDEZ, JM [1 ]
GUIBELALDE, E [1 ]
机构
[1] SAN CARLOS UNIV HOSP,MED PHYS SERV,E-28040 MADRID,SPAIN
关键词
D O I
10.1259/0007-1285-68-815-1215
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Large exposures incurred in interventional radiology procedures make it advisable to establish reference dose values. These dose values should be quoted in quantities representative of the radiological risk to the patient. In Spain, measurement methods were developed to comply with the European Directive on Patient Protection. Dose-area product and, when feasible, surface dose using thermoluminescent dosimetry chips were measured. Both approaches are discussed, as well as their potential use in patient protection programmes. Initial results are presented for a sample of 680 patients in 10 hospital centres in Spain, Mean, median and range are reported for some specific procedures. Mean values of 8750, 6651, 6663, 9292 and 6816 cGy cm(2) are reported for percutaneous transluminal coronary angioplasty, coronary angiography, low extremity, renal and cerebral arteriographies, respectively.
引用
收藏
页码:1215 / 1220
页数:6
相关论文
共 13 条
  • [1] [Anonymous], 1994, MED B, V24, P7
  • [2] A SURVEY OF FLUOROSCOPIC EXPOSURE RATES - AAPM TASK GROUP NO 11 REPORT
    BOONE, JM
    PFEIFFER, DE
    STRAUSS, KJ
    ROSSI, RP
    LIN, PJP
    SHEPARD, JS
    CONWAY, BJ
    [J]. MEDICAL PHYSICS, 1993, 20 (03) : 789 - 794
  • [3] RADIATION-DOSES AND SOMATIC RISK TO PATIENTS DURING CARDIAC RADIOLOGICAL PROCEDURES
    FAULKNER, K
    LOVE, HG
    SWEENEY, JK
    BARDSLEY, RA
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1986, 59 (700) : 359 - 363
  • [4] FAULKNER K, 1989, OPTIMIZATION IMAGE Q, P156
  • [5] HUYSKENS CJ, 1995, RADIAT PROT DOSIM, V57, P475
  • [6] KARPPINEN J, 1995, RADIAT PROT DOSIM, V57, P481
  • [7] PATIENT DOSES RECEIVED DURING DIGITAL SUBTRACTION ANGIOGRAPHY
    STEELE, HR
    TEMPERTON, DH
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (785) : 452 - 456
  • [8] TAYLOR D, 1989, OPTIMIZATION IMAGE Q, P200
  • [9] 1994, STIPUB979 INT AT EN
  • [10] 1994, FEDERAL PERFORMANCE