Haptic Breakage after Transscleral Fixation of a Single-Piece Acrylic Intraocular Lens
被引:4
作者:
Tanaka, Hiroki
[1
,2
]
Fujinami, Kaoru
[1
,2
,3
,4
]
Watanabe, Ken
[1
,2
]
Noda, Toru
[1
,2
]
Akiyama, Kunihiko
[1
,2
]
机构:
[1] Natl Hosp Org, Natl Tokyo Med Ctr, Dept Ophthalmol, Tokyo, Japan
[2] Keio Univ, Natl Tokyo Med Ctr, Natl Inst Sensory Organs, Lab Visual Physiol, Tokyo, Japan
[3] Keio Univ, Sch Med, Dept Ophthalmol, Tokyo, Japan
[4] UCL Inst Ophthalmol, London, England
来源:
CASE REPORTS IN OPHTHALMOLOGY
|
2014年
/
5卷
/
02期
关键词:
Intraocular lens;
Acrylic intraocular lens;
Intraocular lens damage;
Transscleral fixation;
D O I:
10.1159/000365350
中图分类号:
R77 [眼科学];
学科分类号:
100212 ;
摘要:
Purpose: To report the case of a patient with a damaged haptic of an acrylic intraocular lens (IOL) after transscleral IOL fixation. Methods: This is a retrospective and descriptive case report. Results: A 40-year-old man presented with sudden visual loss in the left eye. He had undergone phacoemulsification/aspiration and IOL implantation 3 years previously and IOL repositioning with transscleral fixation 2 years before the initial visit. His visual acuity was 0.3 in the left eye due to corneal edema caused by a foreign body, i.e., the severed haptic tip of the single-piece acrylic IOL, which was surgically removed. The IOL itself was tilted and therefore explanted. The surface of the tip of the haptic suggested that the monofilament suture thread had exerted continuous force across the haptic. Conclusions: The haptics of acrylic IOLs can be damaged after transscleral fixation. (C) 2014 S. Karger AG, Basel
引用
收藏
页码:212 / 216
页数:5
相关论文