CALIBRATION AND APPLICATIONS OF A HIGH-PRECISION PIEZO SCANNER FOR NANOMETROLOGY

被引:1
|
作者
DURSELEN, R [1 ]
GRUNEWALD, U [1 ]
PREUSS, W [1 ]
机构
[1] CARL ZEISS JENA GMBH,DEPT OPT METROL,JENA,GERMANY
关键词
AFM PROFILER; APPLICATIONS; CALIBRATION OF SCANNER; DETERMINATION OF SCANNER LINEARITY; NANOMETROLOGY;
D O I
暂无
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Piezoelectric tube scanners are being used with many scanning probe microscopes for scanning sample surfaces in three dimensions. Since the piezo effect has nonlinearity, hysteresis, and drift, difficulties will occur with distortion, reproducibility, and accuracy of measurements (Barrett and Quate 1991). Correction procedures, which are not based on improvement of linearity of scan movements, will still be affected by remaining nonlinearities of an order of a few percent. The manufacturing of increasing amounts of miniaturized structures for microsystems and semiconductors, as well as certification of standards and measuring devices, will, however, require higher accuracy of profile measurements in the submicrometer and nanometer range. This publication describes a scanner which has a nonlinearity of < 0.05%, a procedure for calibration and detection of nonlinearity, as well as some applications. The calibration technique is based on the fact that manufacturing and measurement of optical diffraction gratings can be very well controlled, allowing light wavelength as an indirect reference in the nanometer range (Antrak 1993, Bartzke et al. 1993, Grunewald 1993, Hutley 1982).
引用
收藏
页码:91 / 96
页数:6
相关论文
共 50 条
  • [21] Linearity calibration method of the high-precision displacement sensor
    Zhang, Defu
    Ge, Chuan
    Li, Xianling
    Ni, Mingyang
    Guo, Kang
    Li, Pengzhi
    Yi Qi Yi Biao Xue Bao/Chinese Journal of Scientific Instrument, 2015, 36 (05): : 982 - 988
  • [22] Calibration of Stellar Luminosity Using High-Precision Parallaxes
    Butkevich, Alexey G.
    CLASSIFICATION AND DISCOVERY IN LARGE ASTRONOMICAL SURVEYS, 2008, 1082 : 66 - 70
  • [23] Image input using an oversize high-precision vertical scanner, and a variety of output applications for the resulting data
    Sakurazawa, M
    DPP 2005: International Conference on Digital Production Printing and Industrial Applications, Final Program and Proceedings, 2005, : 127 - 128
  • [24] Research on a high-precision calibration method for tunable lasers
    Xiang, Na
    Li, Zhengying
    Gui, Xin
    Wang, Fan
    Hou, Yarong
    Wang, Honghai
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2018, 29 (03)
  • [25] High-Precision Calibration of a Weld-On-The-Fly-System
    Emmelmann, Claus
    Schenk, Kerstin
    Wollnack, Joerg
    Kirchhoff, Marc
    LASERS IN MANUFACTURING 2011: PROCEEDINGS OF THE SIXTH INTERNATIONAL WLT CONFERENCE ON LASERS IN MANUFACTURING, VOL 12, PT A, 2011, 12 : 739 - 743
  • [26] Iterative mapping for high-precision calibration and displacement measurements
    Liu, Cong
    Wang, Licheng
    Liu, Xiaopeng
    Xu, Zhihong
    Optik, 2021, 248
  • [27] High-precision camera distortion measurements with a "calibration harp"
    Tang, Zhongwei
    von Gioi, Rafael Grompone
    Monasse, Pascal
    Morel, Jean-Michel
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2012, 29 (10) : 2134 - 2143
  • [28] Calibration errors of high-precision inertial measurement units
    Zhang, Hong-Liang
    Wu, Yuan-Xin
    Zha, Ya-Bing
    Wu, Mei-Ping
    Hu, Xiao-Ping
    Guofang Keji Daxue Xuebao/Journal of National University of Defense Technology, 2010, 32 (01): : 142 - 146
  • [29] High-precision orbit determination and altimeter calibration for ERS
    Zandbergen, R
    Dow, JM
    Fadrique, FMM
    Piriz, R
    12TH INTERNATIONAL SYMPOSIUM ON SPACE FLIGHT DYNAMICS, 1997, 403 : 477 - 484
  • [30] A HIGH-PRECISION CALIBRATION OF THE AD RADIOCARBON TIME SCALE
    STUIVER, M
    RADIOCARBON, 1982, 24 (01) : 1 - 26