Effects of design parameters on the radial force of percutaneous aortic valve stents

被引:10
作者
Kumar, Gideon V. Praveen [1 ]
Mathew, Lazar [1 ]
机构
[1] VIT Univ, Sch Biotechnol Chem & Biomed Engn, Vellore, Tamil Nadu, India
关键词
Stents; Radial force; Finite element analysis; Strut thickness; Design parameters;
D O I
10.1016/j.carrev.2009.04.005
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Valve stents are small cylindrical structures introduced in stenosed valves to replace them and restore blood flow. High radial force has become an essential feature of these stents. The primary aim of this investigation was to investigate the effects of varying stent "design parameters" on its radial force by using the finite element method (FEM). The improved model design was carried out to meet the functional and surgical requirements. The finite element model to evaluate radial force of stent was dependant on various design parameters. The results show that amplitude was the dominant parameter in this study. Increasing either strut circumference or thickness generally improved radial force of stent. However, increasing either amplitude or radius of curvature generally weakened it. In conclusion, FEM can quantify the compressive mechanical property of stent and help designers to optimize stent systems. (C) 2010 Elsevier Inc. All rights reserved.
引用
收藏
页码:101 / 104
页数:4
相关论文
共 7 条
[1]  
Joyy T, 2004, J APPL PHYSIOL, V97, P424
[2]   New stent design for percutaneous aortic valve replacement [J].
Kumar, Gideon V. Praveen ;
Mathew, Lazar .
CARDIOVASCULAR REVASCULARIZATION MEDICINE, 2009, 10 (02) :121-124
[3]  
Lau K. W., 2004, SMJ Singapore Medical Journal, V45, P305
[4]  
McClean D.R., 2002, STENT DESIGN IMPLICA
[5]  
Morton AC, 2004, INFLUENCE PHYS STENT
[6]  
Sangiorgi Giuseppe, 2007, Ann Ist Super Sanita, V43, P89
[7]  
Speed J., 2000, MINIM INVASIV THER, V9, P235