Comparison of intraocular pressure adjusted by central corneal thickness or corneal biomechanical properties as measured in glaucomatous eyes using noncontact tonometers and the Goldmann applanation tonometer

被引:11
|
作者
Yaoeda, Kiyoshi [1 ,2 ]
Fukushima, Atsushi [1 ]
Shirakashi, Motohiro [3 ]
Fukuchi, Takeo [2 ]
机构
[1] Yaoeda Eye Clin, 2-1649-1 Naga Chou, Nagaoka, Niigata 9400053, Japan
[2] Niigata Univ, Grad Sch Med & Dent Sci, Div Ophthalmol & Visual Sci, Niigata, Japan
[3] Kido Eye Clin, Niigata, Japan
来源
CLINICAL OPHTHALMOLOGY | 2016年 / 10卷
基金
日本学术振兴会;
关键词
ocular response analyzer; corneal biomechanical property; corneal hysteresis; glaucoma; intraocular pressure;
D O I
10.2147/OPTH.S106836
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To investigate the correlation coefficients between intraocular pressure (IOP) before and after adjusting for central corneal thickness (CCT) and corneal biomechanical properties. Patients and methods: A total of 218 eyes of 218 patients with primary open-angle glaucoma (mean age =71.5 years; mean spherical equivalent =-0.51 D; mean deviation determined by Humphrey visual field analyzer =-3.22 dB) were included in this study. The tIOP and tIOPCCT, which were adjusted by the CCT (with tIOP meaning IOP not adjusted by CCT, as determined using the CT-1P; and tIOPCCT meaning IOP adjusted by CCT, as determined using the CT-1P), were determined using a noncontact tonometer. The IOPg and IOPCCT, which were adjusted by CCT, and IOPcc adjusted by corneal biomechanical properties were determined using a Reichert 7CR (with IOPg meaning IOP not adjusted by CCT or corneal biomechanical properties, as determined using the Reichert 7CR; IOPCCT meaning IOP adjusted by CCT, as determined using the Reichert 7CR; and IOPcc meaning IOP adjusted by corneal biomechanical properties, as determined using the Reichert 7CR). The GT and GTCCT adjusted by CCT were determined using a Goldmann applanation tonometer (with GT meaning IOP not adjusted by CCT, as determined using the Goldmann applanation tonometer; and with GTCCT meaning IOP adjusted by CCT, as determined using the GAT). Pearson's correlation coefficients among the IOPs were calculated and compared. P-values <0.05 were considered as statistically significant. Results: The tIOP, tIOPCCT, IOPg, IOPCCT, IOPcc, GT, and GTCCT were 14.8 +/- 2.5, 15.0 +/- 2.4, 13.1 +/- 3.2, 13.3 +/- 3.1, 13.7 +/- 2.9, 13.2 +/- 2.4, and 13.4 +/- 2.3 mmHg (mean +/- standard deviation), respectively. The correlation coefficient between tIOPCCT and tIOP (r=0.979) was significantly higher than that between tIOPCCT and the other IOPs (r=0.668-0.852; P<0.001, respectively). The correlation coefficient between IOPCCT and IOPg (r=0.994) or IOPcc and IOPg (r=0.892) was significantly higher than that between IOPCCT or IOPcc and the other IOPs (r=0.669-0.740; P<0.001, respectively). The correlation coefficient between GTCCT and GT (r=0.989) was significantly higher than that between GTCCT and the other IOPs (r=0.669-0.740; P<0.001, respectively). Conclusion: The IOP adjusted by CCT or corneal biomechanical properties depends on the measurement instrument itself, rather than the adjustment methods, for eyes of patients with primary open-angle glaucoma.
引用
收藏
页码:829 / 834
页数:6
相关论文
共 50 条
  • [41] Variations of Intraocular Pressure Measured by Goldmann Applanation Tonometer, Tono-Pen, iCare Rebound Tonometer, and Pascal Dynamic Contour Tonometer in Patients With Corneal Edema After Phacoemulsification
    Kiddee, Weerawat
    Tanjana, Arunee
    JOURNAL OF GLAUCOMA, 2021, 30 (04) : 317 - 324
  • [42] Clinical comparison of ocular blood flow tonometer, tonopen, and Goldmann applanation tonometer for measuring intraocular pressure in postkeratoplasty eyes
    Rao, VJ
    Gnanaraj, L
    Mitchell, KW
    Figueiredo, FC
    CORNEA, 2001, 20 (08) : 834 - 838
  • [43] Assessment of intraocular pressure measurement between Goldman applanation tonometer, rebound tonometer, non-contact tonometer, and its correlation with central corneal thickness
    Porwal, Amit C.
    Shrishrimal, Meghna
    Punamia, Ravin P.
    Mathew, Bennet Chacko
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2023, 71 (05) : 1927 - 1931
  • [44] A comparison of intraocular pressure values obtained using a Goldmann applanation tonometer and a handheld version of applanation resonance tonometer: A preliminary report
    Mulak, Malgorzata
    Czak, Wojciech A.
    Mimier, Malgorzata
    Kaczmarek, Radoslaw
    ADVANCES IN CLINICAL AND EXPERIMENTAL MEDICINE, 2018, 27 (04): : 481 - 485
  • [45] Comparison of Goldmann applanation tonometry and Pascal dynamic contour tonometry in relation to central corneal thickness and corneal curvature
    Andreanos, Konstantinos
    Koutsandrea, Chryssanthi
    Papaconstantinou, Dimitris
    Diagourtas, Andreas
    Kotoulas, Andreas
    Dimitrakas, Panagiotis
    Moschos, Marilita M.
    CLINICAL OPHTHALMOLOGY, 2016, 10 : 2477 - 2484
  • [46] Comparison of Intraocular Pressure Measurements With the Reichert Pt100, the Keeler Pulsair Intellipuff Portable Noncontact Tonometers, and Goldmann Applanation Tonometry
    Hubanova, Ralitsa
    Aptel, Florent
    Zhou, Thierry
    Arnol, Nathalie
    Romanet, Jean-Paul
    Chiquet, Christophe
    JOURNAL OF GLAUCOMA, 2015, 24 (05) : 356 - 363
  • [47] Effect of corneal thickness on intraocular pressure measurements with the Pascal dynamic contour, Canon TX-10 non-contact and Goldmann applanation tonometers in healthy subjects
    Pelit, Aysel
    Altan-Yaycioglu, Rana
    Pelit, Aykut
    Akova, Yonca A.
    CLINICAL AND EXPERIMENTAL OPTOMETRY, 2009, 92 (01) : 14 - 18
  • [48] Cross-Sectional Study of Differences between Intraocular Pressure Measurements using Goldmann, iCare, and Air-Puff Tonometers and their Correlation with Central Corneal Thickness
    Mandour, Sameh Saad
    Elframawy, Ahmed
    Murad, Mohammad Moataz
    Nage, Sara Abd Elmegeed
    JOURNAL OF CURRENT OPHTHALMOLOGY, 2023, 35 (04):
  • [49] The applicability of correction factor for corneal thickness on non-contact tonometer measured intraocular pressure in LASIK treated eyes
    Jethani, Jitendra
    Dave, Paaraj
    Jethani, Monica
    Desai, Yogesh
    Patel, Purvi
    SAUDI JOURNAL OF OPHTHALMOLOGY, 2016, 30 (01) : 25 - 28
  • [50] Comparison of Intraocular Pressure Measurements Obtained by Icare PRO Rebound Tonometer, Tomey FT-1000 Noncontact Tonometer, and Goldmann Applanation Tonometer in Healthy Subjects
    Guler, Mete
    Bilak, Semsettin
    Bilgin, Burak
    Simsek, Ali
    Capkin, Musa
    Reyhan, Ali Hakim
    JOURNAL OF GLAUCOMA, 2015, 24 (08) : 613 - 618