PHOSPHATE KINETICS DURING HIGH-FLUX HEMODIALYSIS

被引:0
作者
DESOI, CA [1 ]
UMANS, JG [1 ]
机构
[1] UNIV CHICAGO, DEPT MED,NEPHROL SECT,5841 S MARYLAND AVE,MC-5100, CHICAGO, IL 60637 USA
来源
JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY | 1993年 / 4卷 / 05期
关键词
PHOSPHORUS; CLEARANCE; HEMODIALYSIS;
D O I
暂无
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Phosphate clearance by polysulfone (PS) and cuprophane (CU) membranes and the relationship of peridialytic changes in serum phosphate with those of serum-ionized calcium, parathyroid hormone (PTH), and insulin were studied in six stable patients undergoing chronic hemodialysis (HD). Dietary phosphate intake was 25.7 mmol/day, and total dose of elemental calcium was 3.2 g/day. Patients were dialyzed for 2 to 4 h, once with each membrane. Serum phosphate levels fell precipitously during the first hour of HD with both dialyzers, from 1.42 and 1.49 mM to nadirs of 0.53 and 0.69 mM for PS and CU, respectively. Phosphate levels began to increase either late in HD or at the end of HD and, by 4 h post-HD, did not differ from predialysis values. Total mass transfer of phosphate was greater during the first hour (3.4 and 3.2 mmol for PS and CU, respectively) than during the remainder of HD (2.4 and 2.6 mmol/h). There were no significant differences in intradialytic serum phosphate changes, postdialytic phosphate rebound, or total phosphate removal between the two dialyzers. Ionized calcium increased by 0.11 mM, and PTH was suppressed to 40 to 50% of baseline values during dialysis with either membrane. Although phosphate removal continued for the duration of dialysis, serum phosphate did not continue to decrease, either reaching an apparent steady state or beginning to rebound, even during dialysis. This suggests active phosphate mobilization from a pool other than the extracellular fluid and demonstrates the inadequacy of a one-compartment model to explain these data. Further, these data do not support the regulation of intradialytic phosphate mobilization by serum PTH or insulin. The transient, but profound, intradialytic hypophosphatemia may serve as a stimulus for as-yet-uncharacterized homeostatic mechanisms.
引用
收藏
页码:1214 / 1218
页数:5
相关论文
共 50 条
[41]   EFFECT OF HIGH-FLUX HEMODIALYSIS ON QUALITY-OF-LIFE AND NEUROPSYCHOLOGICAL FUNCTION IN CHRONIC-HEMODIALYSIS PATIENTS [J].
CHURCHILL, DN ;
BIRD, DR ;
TAYLOR, DW ;
BEECROFT, ML ;
GORMAN, J ;
WALLACE, JE .
AMERICAN JOURNAL OF NEPHROLOGY, 1992, 12 (06) :412-418
[42]   Phosphate Kinetics During Weekly Cycle of Hemodialysis Sessions: Application of Mathematical Modeling [J].
Debowska, Malgorzata ;
Poleszczuk, Jan ;
Wojcik-Zaluska, Alicja ;
Ksiazek, Andrzej ;
Zaluska, Wojciech .
ARTIFICIAL ORGANS, 2015, 39 (12) :1005-1014
[43]   Removal of vancomycin administered during dialysis by a high-flux dialyzer [J].
Nyman, Heather A. ;
Agarwal, Adhish ;
Senekjian, Harry O. ;
Leypoldt, John K. ;
Cheung, Alfred K. .
HEMODIALYSIS INTERNATIONAL, 2018, 22 (03) :383-387
[44]   HIGH-FLUX SYNTHETIC VERSUS CELLULOSIC MEMBRANES FOR BETA-2-MICROGLOBULIN REMOVAL DURING HEMODIALYSIS, HEMODIAFILTRATION AND HEMOFILTRATION [J].
FLOEGE, J ;
GRANOLLERAS, C ;
DESCHODT, G ;
HECK, M ;
BAUDIN, G ;
BRANGER, B ;
TOURNIER, O ;
REINHARD, B ;
EISENBACH, GM ;
SMEBY, LC ;
KOCH, KM ;
SHALDON, S .
NEPHROLOGY DIALYSIS TRANSPLANTATION, 1989, 4 (07) :653-657
[45]   HIGH-FLUX DIALYSIS MEMBRANES IMPROVE LIPID PROFILE IN CHRONIC-HEMODIALYSIS PATIENTS [J].
BLANKESTIJN, PJ ;
VOS, PF ;
RABELINK, TJ ;
VANRIJN, HJM ;
JANSEN, H ;
KOOMANS, HA .
JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 1995, 5 (09) :1703-1708
[46]   Use of Charcoal Hemoperfusion and High-Flux Hemodialysis in Carbamazepine Intoxication: Two Case Reports [J].
Kocak, Sibel Yucel ;
Kudu, Aysegul ;
Gumusel, Ceren ;
Kayalar, Arzu Ozdemir ;
Yilmaz, Murvet .
IRANIAN RED CRESCENT MEDICAL JOURNAL, 2019, 21 (12)
[47]   Administration and clearance of amphotericin B during high-efficiency or high-efficiency/high-flux dialysis [J].
Gussak, HM ;
Rahman, S ;
Bastani, B .
AMERICAN JOURNAL OF KIDNEY DISEASES, 2001, 37 (06) :art. no.-e45
[48]   Role of the dialyzer membrane on the overall phosphate kinetics during hemodialysis [J].
Katopodis, KP ;
Chala, A ;
Koliousi, E ;
Takouli, L ;
Kalaitzidis, R ;
Theodorou, J ;
Siamopoulos, KC .
BLOOD PURIFICATION, 2005, 23 (05) :359-364
[49]   Comparison between On-Line High-Efficiency Hemodiafiltration and Conventional High-Flux Hemodialysis for Polyclonal Free Light Chain Removal [J].
Lamy, Thomas ;
Henri, Patrick ;
Lobbedez, Thierry ;
Comby, Elisabeth ;
Ryckelynck, Jean-Philippe ;
Ficheux, Maxence .
BLOOD PURIFICATION, 2014, 37 (02) :93-98
[50]   Multiple-dose pharmacokinetics of cefepime in long-term hemodialysis with high-flux membranes [J].
S. Schmaldienst ;
F. Traunmüller ;
H. Burgmann ;
A. R. Rosenkranz ;
R. Thalhammer-Scherrer ;
W. H. Hörl ;
F. Thalhammer .
European Journal of Clinical Pharmacology, 2000, 56 :61-64